Part 1.
DEPARTMENT OF STATE HEALTH SERVICES
Chapter 169.
ZOONOSIS CONTROL
Subchapter F. REPTILE-ASSOCIATED SALMONELLOSIS
25 TAC §169.121
The Executive Commissioner of the Health and Human Services
Commission, on behalf of the Department of State Health Services (department),
proposes an amendment to §169.121, concerning warnings retail pet stores
must provide relating to reptile-associated salmonellosis.
BACKGROUND AND PURPOSE
This rule is necessary to comply with Health and Safety Code, Chapter 81,
Subchapter I, "Animal-Borne Diseases," which requires retail pet stores to
post signs and distribute warnings relating to reptile-associated salmonellosis
to purchasers of reptiles. The signs and warnings are to be in accordance
with the form and content designated by the department.
Government Code, §2001.039, requires that each state agency review
and consider for re-adoption each rule adopted by that agency pursuant to
the Government Code, Chapter 2001 (Administrative Procedure Act). Section
169.121 has been reviewed and the department has determined that reasons for
adopting the section continue to exist because a rule on this subject is needed.
After carefully considering the alternatives, the department believes the
rule as amended is the best method of implementing the statute, and informing,
without unduly alarming, the pet-buying public.
SECTION-BY-SECTION SUMMARY
The amendment to §169.121 is necessary to comply with the mandated
four-year rule review, update the legacy agency name, and make the guidelines
for warning signs consistent with the Centers for Disease Control and Prevention
(CDC) recommendations.
FISCAL NOTE
Martha McGlothlin, Section Director, Community Preparedness Section, has
determined that for each year of the first five-year period that the section
will be in effect, there will be no fiscal implications to state or local
governments as a result of enforcing and administering the section as proposed.
SMALL AND MICRO-BUSINESS IMPACT ANALYSIS
Ms. McGlothlin has also determined that there will be no effect on small
businesses or micro-businesses required to comply with the section as proposed.
This was determined by interpretation of the rule that small businesses and
micro-businesses will not be required to alter their business practices in
order to comply with the section. There are no anticipated economic costs
to persons who are required to comply with the section as proposed. There
is no anticipated negative impact on local employment.
PUBLIC BENEFIT
In addition, Ms. McGlothlin has also determined that for each year of the
first five years the section is in effect, the public will benefit from adoption
of the section. The public benefit anticipated as a result of enforcing or
administering the section will be the increased public awareness of the risk
involved with having reptiles as pets, as it pertains to reptile-associated
salmonellosis.
REGULATORY ANALYSIS
The department has determined that this proposal is not a "major environmental
rule" as defined by Government Code, §2001.0225. "Major environmental
rule" is defined to mean a rule the specific intent of which is to protect
the environment or reduce risk to human health from environmental exposure
and that may adversely affect, in a material way, the economy, a sector of
the economy, productivity, competition, jobs, the environment or the public
health and safety of a state or a sector of the state.
TAKINGS IMPACT ASSESSMENT
The department has determined that the proposed amendment does not restrict
or limit an owner's right to his or her property that would otherwise exist
in the absence of government action and, therefore, does not constitute a
taking under Government Code, §2007.043.
PUBLIC COMMENT
Comments on the proposal may be submitted to Tom Sidwa, DVM, Department
of State Health Services, Community Preparedness Section, Zoonosis Control
Branch, 1100 West 49th Street, Austin, Texas 78756, or by email to Tom.Sidwa@dshs.state.tx.us.
Comments will be accepted for 30 days following publication of the proposal
in the
Texas Register
.
LEGAL CERTIFICATION
The Department of State Health Services General Counsel, Cathy Campbell,
certifies that the proposed rule has been reviewed by legal counsel and found
to be within the state agencies' authority to adopt.
STATUTORY AUTHORITY
The proposed amendment is authorized by Health and Safety Code, §81.352,
which requires the department to adopt a rule governing the form and content
of the sign and written warning relating to reptile-associated salmonellosis;
and Government Code, §531.0055, and Health and Safety Code, §1001.075,
which authorize the Executive Commissioner of the Health and Human Services
Commission to adopt rules and policies necessary for the operation and provision
of health and human services by the department and for the administration
of Health and Safety Code, Chapter 1001.
The proposed amendment affects Health and Safety Code, Chapters 81 and
1001; and Government Code, Chapter 531. Review of the rule implements Government
Code, §2001.039.
§169.121.Reptile-Associated Salmonellosis.
(a)
The Health and Safety Code, §81.352, requires retail
stores that sell reptiles to post warning signs and distribute written warnings
regarding reptile-associated salmonellosis to purchasers in accordance with
the form and content designated by the
Department of State Health Services
[
(b)
The warning signs must meet the following guidelines.
(1) - (2)
(No change.)
(3)
At a minimum, the contents of the sign must include the
following recommendations for preventing transmission of
Salmonella
from reptiles to humans.
(A)
(No change.)
(B)
Persons at increased risk for infection or serious complications
of salmonellosis (e.g., children aged <5 years and immunocompromised persons)
should avoid contact with reptiles
and any items that have been in contact
with reptiles
.
(C)
Reptiles
[
(D)
Reptiles
[
(E)
Reptiles
[
(F)
Reptiles
[
(4)
(No change.)
(c)
(No change.)
This agency hereby certifies that the proposal has been reviewed
by legal counsel and found to be within the agency's legal authority to adopt.
Filed with the Office of
the Secretary of State on June 6, 2006.
TRD-200603055
Cathy Campbell
General Counsel
Department of State Health Services
Earliest possible date of adoption: July 23, 2006
For further information, please call: (512) 458-7111 x6972
25 TAC §169.131
The Executive Commissioner of the Health and Human Services
Commission, on behalf of the Department of State Health Services (department),
proposes an amendment to §169.131, concerning caging requirements and
standards for the keeping and confinement of dangerous wild animals.
BACKGROUND AND PURPOSE
This rule is necessary to comply with Health and Safety Code, Chapter 822,
Subchapter E, "Dangerous Wild Animals," which requires owners of a dangerous
wild animal to keep and confine the animal in accordance with the caging requirements
and standards established by the department.
Government Code, §2001.039, requires that each state agency review
and consider for re-adoption each rule adopted by that agency pursuant to
the Government Code, Chapter 2001 (Administrative Procedure Act). Section
169.131 has been reviewed and the department has determined that reasons for
adopting the section continue to exist because a rule on this subject is needed.
After carefully considering the alternatives, the department believes the
rule as amended is the best method of implementing the statute, providing
for enclosures that protect public safety and provide a safe and secure enclosure
for the animal; without undue burden on owners.
SECTION-BY-SECTION SUMMARY
The amendment to §169.131 is necessary to comply with the mandated
four-year rule review and to clarify the minimum caging requirements relating
to the structures and outdoor facilities containing dangerous wild animals.
FISCAL NOTE
Martha McGlothlin, Section Director, Community Preparedness Section, has
determined that for each year of the first five-year period that the section
will be in effect, there will be no fiscal implications to state or local
governments as a result of enforcing and administering the section as proposed.
SMALL AND MICRO-BUSINESS IMPACT ANALYSIS
Ms. McGlothlin has also determined that there will be no effect on small
businesses or micro-businesses required to comply with the section as proposed.
This was determined by interpretation of the rule that small businesses and
micro-businesses will not be required to alter their business practices in
order to comply with the section. There are no anticipated economic costs
to persons who are required to comply with the section as proposed. There
is no anticipated negative impact on local employment.
PUBLIC BENEFIT
In addition, Ms. McGlothlin has also determined that for each year of the
first five years the section is in effect, the public will benefit from adoption
of the section. The public benefit anticipated as a result of enforcing or
administering the section will be that it enhances public health and safety
by keeping dangerous wild animals contained in safe, healthy, and humane environments.
REGULATORY ANALYSIS
The department has determined that this proposal is not a "major environmental
rule" as defined by Government Code, §2001.0225. "Major environmental
rule" is defined to mean a rule the specific intent of which is to protect
the environment or reduce risk to human health from environmental exposure
and that may adversely affect, in a material way, the economy, a sector of
the economy, productivity, competition, jobs, the environment or the public
health and safety of a state or a sector of the state. This proposal is not
specifically intended to protect the environment or reduce risks to human
health from environmental exposure.
TAKINGS IMPACT ASSESSMENT
The department has determined that the proposed amendment does not restrict
or limit an owner's right to his or her property that would otherwise exist
in the absence of government action and, therefore, does not constitute a
taking under Government Code, §2007.043.
PUBLIC COMMENT
Comments on the proposal may be submitted to Tom Sidwa, DVM, Department
of State Health Services, Community Preparedness Section, Zoonosis Control
Branch, 1100 West 49th Street, Austin, Texas 78756, or by email to Tom.Sidwa@dshs.state.tx.us.
Comments will be accepted for 30 days following publication of the proposal
in the
Texas Register
.
LEGAL CERTIFICATION
The Department of State Health Services General Counsel, Cathy Campbell,
certifies that the proposed rule has been reviewed by legal counsel and found
to be within the state agencies' authority to adopt.
STATUTORY AUTHORITY
The proposed amendment is authorized by Health and Safety Code, §822.111,
which requires the department to establish the caging requirements and standards
for the keeping and confinement of dangerous wild animals; and Government
Code, §531.0055, and Health and Safety Code, §1001.075, which authorize
the Executive Commissioner of the Health and Human Services Commission to
adopt rules and policies necessary for the operation and provision of health
and human services by the department and for the administration of Health
and Safety Code, Chapter 1001.
The proposed amendment affects Health and Safety Code, Chapters 822 and
1001; and Government Code, Chapter 531. Review of the rule implements Government
Code, §2001.039.
§169.131.Caging Requirements and Standards for Dangerous Wild Animals.
(a)
Definitions.
(1)
Key components of facilities for confining dangerous wild
animals and restricting public contact with the animals are the primary enclosure
and the perimeter fence.
(A)
Primary enclosure - Any structure used to immediately restrict
an
animal(s)
[
(B)
(No change.)
(2)
Where specified in this section, primary enclosures for
dangerous wild animals shall be equipped to provide for
a safe, healthy,
and humane environment for
[
(A)
(No change.)
(B)
Shelter, nest box, or den - A structure that protects the
animal(s)
[
(i)
Shelter - A structure that provides protection from the
elements and from extremes in temperature that are detrimental to the health
and welfare of the
animal(s)
[
(ii)
(No change.)
(C) - (D)
(No change.)
(b)
General Requirements.
(1)
Primary enclosures for housing dangerous wild animals shall
be sufficiently strong to prevent escape and to protect the
animal(s)
[
(2)
(No change.)
(3)
A perimeter fence, sufficient to deter entry by the public,
shall be a minimum of 8 feet in height and shall completely surround the premises
where
the animal(s) is
[
(c)
Structural Requirements for Primary Enclosures. In addition
to the size and equipment requirements for primary enclosures, dangerous wild
animals shall be caged in accordance with the following requirements.
(1)
All primary enclosures shall be equipped with a safety
entrance. Such entrances shall include a double-door mechanism, interconnecting
cages, a lock-down area, or other comparable devices that will prevent escape
and safeguard the keeper. Safety entrances shall be constructed of materials
that are of equivalent strength
or greater than
[
(2)
All primary enclosures constructed of chain link or other
approved materials shall be well braced and securely anchored at or below
ground level to prevent escape by digging or erosion. Metal clamps, ties,
or braces used in the construction of enclosures shall be of strength equivalent
to
or greater than
the material required for primary enclosure
construction for the particular species.
(3)
Additional minimum requirements for specific species and
hybrids of those species shall be as follows.
(A)
Chimpanzees, gorillas, and orangutans.
(i)
Outdoor facilities - Construction material shall consist
of steel bars, 2-inch galvanized pipe, masonry block, or their strength equivalent
or greater
.
(ii)
Indoor facilities - Potential escape routes shall be equipped
, at minimum,
with steel bars, 2-inch galvanized pipe, or equivalent.
(B) - (C)
(No change.)
(d)
Primary Enclosure Size and Equipment Requirements. No dangerous
wild animal shall be confined in any primary enclosure that contains more
individual animals than specified in this section, is smaller in dimension
than specified in this section, or is not equipped as specified in this section.
The area occupied by pools, ponds, or lakes shall be in addition to the space
requirements for the primary enclosure.
(1)
Primates.
(A) - (C)
(No change.)
(D)
Requirements for specific primate species are as follows:
(i)
Baboons. For one animal the primary enclosure shall have
a
minimum
floor area of 100 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size
shall be increased by
at least
100 square feet.
(ii)
Chimpanzees. For one animal the primary enclosure shall
have a
minimum
floor area of 200 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size
shall be increased by
at least
100 square feet.
(iii)
Orangutans. For one animal the primary enclosure shall
have a
minimum
floor area of 200 square feet with a wall or fence
at least
10 feet high. For each additional animal primary enclosure
size shall be increased by
at least
200 square feet.
(iv)
Gorillas. For one animal the primary enclosure shall have
a
minimum
floor area of 300 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size
shall be increased by
at least
200 square feet.
(2)
Wild felines.
(A)
(No change.)
(B)
Each primary enclosure shall have [
(C)
Each primary enclosure shall have an elevated
platform(s)
[
(D)
Each primary enclosure shall have
at least one
[
(E)
Requirements for specific species of wild felines are as
follows:
(i)
Lions, tigers, and cheetahs.
(I)
For one animal the primary enclosure shall have a
minimum
floor area of 300 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size shall
be increased by
at least
150 square feet.
(II)
Outdoor primary enclosures over 1,000 square feet (uncovered)
shall have vertical jump walls at least 10 feet high with a 45 degree inward
angle overhang
at least
2 feet wide or jump walls at least 12 feet
high without an overhang. The inward angle fencing shall be made of the same
material as the vertical fencing.
(ii)
Jaguars, leopards, and cougars.
(I)
For one animal the primary enclosure shall have a
minimum
floor area of 200 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size shall
be increased by
at least
100 square feet.
(II)
(No change.)
(iii)
Bobcats, lynxes, ocelots, caracals, and servals. For
one animal the primary enclosure shall have a
minimum
floor area
of 80 square feet with a wall or fence
at least
8 feet high. For
each additional animal primary enclosure size shall be increased by
at least
40 square feet.
(3)
Bears.
(A)
(No change.)
(B)
Each primary enclosure shall have [
(C)
Each primary enclosure shall have an elevated platform(s)
for resting
that will accommodate all animals in the enclosure simultaneously
.
(D)
Requirement for specific types of bears are as follows:
(i)
Sun bears.
(I)
For one animal the primary enclosure shall have a
minimum
floor area of 200 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size shall
be increased by
at least
100 square feet.
(II)
(No change.)
(ii)
Black bears and Asiatic bears.
(I)
For one animal the primary enclosure shall have a
minimum
floor area of 300 square feet with a wall or fence
at least
8 feet high. For each additional animal primary enclosure size shall
be increased by
at least
150 square feet.
(II)
(No change.)
(iii)
Brown bears and polar bears.
(I)
For one animal the primary enclosure shall have a
minimum
floor area of 400 square feet with a wall or fence
at least
10 feet high. For each additional animal primary enclosure size shall
be increased by
at least
200 square feet.
(II) - (III)
(No change.)
(4)
Coyotes and jackals.
(A)
(No change.)
(B)
Each primary enclosure shall have [
(C)
For one animal the primary enclosure shall have a
minimum
floor area of 150 square feet with a wall or fence
at least
6 feet high. For each additional animal primary enclosure size shall
be increased by
at least
100 square feet.
(D)
Each primary enclosure shall have an elevated
platform(s)
[
(E)
Uncovered outdoor primary enclosures over 1,000 square
feet shall have vertical jump walls at least 8 feet high with a 45 degree
inward angle overhang
at least
2 feet wide or jump walls
at least
10 feet high without an overhang. The inward angle fencing
shall be made of the same material as the vertical fencing.
(5)
Hyenas.
(A)
For one animal the primary enclosure shall have a
minimum
floor area of 200 square feet with a wall or fence
at least
6 feet high. For each additional animal primary enclosure size shall
be increased by
at least
100 square feet.
(B)
Each primary enclosure shall have an elevated
platform(s)
[
(C)
Outdoor primary enclosures over 1,000 square feet (uncovered)
shall have vertical jump walls at least 8 feet high with a 45 degree inward
angle overhang
at least
2 feet wide or jump walls
at least
10 feet high without an overhang. The inward angle fencing shall be
made of the same material as the vertical fencing.
This agency hereby certifies that the proposal
has been reviewed by legal counsel and found to be within the agency's legal
authority to adopt.
Filed with the Office of
the Secretary of State on June 6, 2006.
TRD-200603054
Cathy Campbell
General Counsel
Department of State Health Services
Earliest possible date of adoption: July 23, 2006
For further information, please call: (512) 458-7111 x6972
Subchapter E. REGISTRATION REGULATIONS
The Executive Commissioner of the Health and Human Services Commission
(commission), on behalf of the Department of State Health Services (department),
proposes the repeal of §289.230, concerning certification of mammography
systems and accreditation of mammography facilities and new §289.230,
concerning certification of mammography systems and mammography machines used
for interventional breast radiography.
BACKGROUND AND PURPOSE
Section 289.230 concerning certification of mammography systems and accreditation
of mammography facilities is being repealed and divided into two new rules,
one concerning certification and the other concerning accreditation. The repeal
and new rules are the result of the agency applying to the United States Food
and Drug Administration (FDA) to become a certifying body for mammography
facilities. The FDA recommended that the current rules be separated into certification
and accreditation rules for clarification, as all mammography facilities in
the state must have certification with the agency, while accreditation may
be with the agency or with the American College of Radiology. Certification
requirements will be located in §289.230, "Certification of Mammography
Systems and Mammography Machines Used for Interventional Breast Radiography."
The accreditation requirements will be incorporated into new §289.234,
"Mammography Accreditation," that is addressed in a separate preamble.
Government Code, §2001.039, requires that each state agency review
and consider for readoption each rule adopted by that agency pursuant to the
Government Code, Chapter 2001 (Administrative Procedure Act). Section 289.230
has been reviewed and the department has determined that reasons for adopting
the section continue to exist because rules on this subject are needed.
SECTION-BY-SECTION SUMMARY
The new section adds definitions for agency accreditation body, agency
certifying body, practitioner of the healing arts, and provisional certificate
to define terms used in this section. Definitions for certification, mean
optical density, medical physicist, and targeted clinical image review are
revised to be consistent with the FDA's Mammography Quality Standards Act
(MQSA) rules. Requirements for certification that are consistent with FDA
certification are added. Adding these requirements will allow the agency to
become a certifying body for mammography facilities. Requirements for suspension
or revocation of facility certification, denials of certification, and repeals
are included. Requirements for machines used for interventional breast radiography
are revised. There is clarification that facilities utilizing physicians and
technologists from temporary agencies must meet qualification standards. Language
on holding patients or image receptors during an exam and use of protective
devices are added. Record-keeping requirements are revised to be consistent
with those of the FDA.
FISCAL NOTE
Susan E. Tennyson, Section Director, Environmental and Consumer Safety
Section, has determined that for each year of the first five-year period that
the sections will be in effect, there will be no fiscal implications to state
or local governments as a result of enforcing and administering the sections
as proposed.
SMALL AND MICRO-BUSINESS IMPACT ANALYSIS
Ms. Tennyson has also determined that there will be no effect on small
businesses or micro-businesses required to comply with the sections as proposed.
This was determined by interpretation of the rule that small businesses and
micro-businesses will not be required to alter their business practices in
order to comply with the sections. There is no anticipated negative impact
on local employment.
PUBLIC BENEFIT
In addition, Ms. Tennyson has also determined that for each year of the
first five years the sections are in effect, the public will benefit from
adoption of the sections. The public benefit anticipated as a result of enforcing
or administering the sections is to protect public health and safety by affording
facilities the opportunity to become certified and accredited for mammography
through one agency. Facilities will also have the option of becoming accredited
through the American College of Radiology.
REGULATORY ANALYSIS
The department has determined that this proposal is not a "major environmental
rule" as defined by Government Code, §2001.0225. "Major environmental
rule" is defined to mean a rule the specific intent of which is to protect
the environment or reduce risk to human health from environmental exposure
and that may adversely affect, in a material way, the economy, a sector of
the economy, productivity, competition, jobs, the environment or the public
health and safety of a state or a sector of the state. This proposal is not
specifically intended to protect the environment or reduce risks to human
health from environmental exposure.
TAKINGS IMPACT ASSESSMENT
The department has determined that the proposed rules do not restrict or
limit an owner's right to his or her property that would otherwise exist in
the absence of government action and, therefore, does not constitute a taking
under Government Code, §2007.043.
PUBLIC COMMENT
Comments on the proposal may be submitted to Cindy Cardwell, Radiation
Group, Policy/Standards/Quality Assurance Unit, Environmental and Consumer
Safety Section, Division of Regulatory Services, Department of State Health
Services, 1100 West 49th Street, Austin, Texas 78756, (512) 834-6770, extension
2239, or by email to cindy.cardwell@dshs.state.tx.us. Comments will be accepted
for 30 days following publication of the proposal in the
Texas Register
.
PUBLIC HEARING
A public hearing to receive comments on the proposal will be scheduled
after publication in the
Texas Register
and
will be held at the Department of State Health Services, Exchange Building,
8407 Wall Street, Austin, Texas 78754. The meeting date will be posted on
the Radiation Control web site (www.dshs.state.tx.us/radiation). Please contact
Cindy Cardwell at (512) 834-6770, extension 2239, or cindy.cardwell@dshs.state.tx.us,
if you have questions.
LEGAL CERTIFICATION
The Department of State Health Services General Counsel, Cathy Campbell,
certifies that the proposed rules have been reviewed by legal counsel and
found to be within the state agencies' authority to adopt.
25 TAC §289.230
(Editor's note: The text of the following section proposed for
repeal will not be published. The section may be examined in the offices of
the Department of State Health Services or in the Texas Register office, Room
245, James Earl Rudder Building, 1019 Brazos Street, Austin.)
STATUTORY AUTHORITY
The proposed repeal is authorized by Health and Safety Code, §401.051,
which provides the Executive Commissioner of the Health and Human Services
Commission with authority to adopt rules and guidelines relating to the control
of radiation, and Government Code, §531.0055, and Health and Safety Code, §1001.075,
which authorize the Executive Commissioner of the Health and Human Services
Commission to adopt rules and policies necessary for the operation and provision
of health and human services by the department and for the administration
of Health and Safety Code, Chapter 1001.
The repeal affects the Health and Safety Code, Chapter 401 and 1001; and
Government Code, Chapter 531.
§289.230.Certification of Mammography Systems and Accreditation of Mammography Facilities.
This agency hereby certifies that the proposal has been
reviewed by legal counsel and found to be within the agency's legal authority
to adopt.
Filed with the Office of
the Secretary of State on June 9, 2006.
TRD-200603126
Cathy Campbell
General Counsel
Department of State Health Services
Earliest possible date of adoption: July 23, 2006
For further information, please call: (512) 458-7111 x6972
25 TAC §289.230
STATUTORY AUTHORITY
The proposed new section is authorized by Health and Safety Code, §401.051,
which provides the Executive Commissioner of the Health and Human Services
Commission with authority to adopt rules and guidelines relating to the control
of radiation, and Government Code, §531.0055, and Health and Safety Code, §1001.075,
which authorize the Executive Commissioner of the Health and Human Services
Commission to adopt rules and policies necessary for the operation and provision
of health and human services by the department and for the administration
of Health and Safety Code, Chapter 1001.
The new section affects the Health and Safety Code, Chapter 401 and 1001;
and Government Code, Chapter 531.
§289.230.Certification of Mammography Systems and Mammography Machines Used for Interventional Breast Radiography.
(a)
Purpose.
(1)
This section provides for the certification of mammography
systems and mammography machines used for interventional breast radiography.
No person shall use radiation machines for mammography of humans or for interventional
breast radiography except as authorized in a certification issued by the agency
in accordance with the requirements of this section. Certification by this
agency includes certification of mammography systems and facilities that have
received accreditation by the agency accreditation body or by another United
States Food and Drug Administration (FDA)-approved accreditation body and
certification of mammography machines used for interventional breast radiography.
(2)
The use of all mammography machines certified in accordance
with this section shall be by or under the supervision of a physician licensed
by the Texas Medical Board.
(b)
Scope.
(1)
In addition to the requirements of this section, all registrants
are subject to the requirements of §289.203 of this title (relating to
Notices, Instructions, and Reports to Workers; Inspections), §289.204
of this title (relating to Fees for Certificates of Registration, Radioactive
Material Licenses, Emergency Planning and Implementation, and Other Regulatory
Services), §289.205 of this title (relating to Hearing and Enforcement
Procedures), §289.226 of this title (relating to Registration of Radiation
Machine Use and Services), and §289.231 of this title (relating to General
Provisions and Standards for Protection Against Machine-Produced Radiation).
Mammography facilities choosing to be accredited by the agency accreditation
body will be subject to §289.234 of this title (relating to Mammography
Accreditation).
(2)
The procedures found in §289.205 of this title for
modifications, suspensions, revocations, denials, and hearings regarding certificates
of registration are applicable to certifications issued by the agency.
(3)
This section does not apply to an entity under the jurisdiction
of the federal government.
(c)
Definitions. The following words and terms, when used in
this section, shall have the following meanings unless the context clearly
indicates otherwise.
(1)
Accreditation--An approval of a mammography machine within
a mammography facility by an accreditation body. A facility may be accredited
by the agency accreditation body or another FDA-approved accreditation body.
(2)
Act--Texas Radiation Control Act, Health and Safety Code,
Chapter 401.
(3)
Action limit--The minimum or maximum value of a quality
assurance measurement representing acceptable performance. Values less than
the minimum or greater than the maximum action limit indicate that corrective
action must be taken by the facility.
(4)
Additional mammography review (includes targeted clinical
image reviews)--At the request of the agency certification body or an FDA-approved
accreditation body, a review by the FDA-approved accreditation body of clinical
images and other relevant facility information necessary to assess conformation
with the accreditation standards. The reviews include the following:
(A)
clinical image review with interpretation; or
(B)
clinical image review without interpretation.
(5)
Adverse event--An undesirable experience associated with
mammography activities within the scope of this section. Adverse events include
but are not limited to:
(A)
poor image quality;
(B)
failure to send mammography reports within 30 days to the
referring physician or in a timely manner to the self-referred patient; and
(C)
use of personnel who do not meet the applicable requirements
of subsection (r) of this section.
(6)
Agency accreditation body--For the purpose of this section,
the agency as approved by the FDA under Title 21, Code of Federal Regulations
(CFR), Part 900.3(d) to accredit mammography facilities in the State of Texas.
(7)
Agency certifying body--For the purpose of this section,
the agency, as approved by FDA, under Title 21, CFR, Part 900.21, to certify
facilities within the State of Texas to perform mammography services.
(8)
Air kerma--The kerma in a given mass of air. The unit used
to measure the quantity of air kerma is the Gray (Gy). For x-rays with energies
less than 300 kiloelectronvolts (keV), 1 Gy = 100 rad. In air, 1 Gy of absorbed
dose is delivered by 114 roentgens (R) of exposure.
(9)
Automatic exposure control (AEC)--A device that automatically
controls one or more technique factors in order to obtain at preselected locations
a required quantity of radiation.
(10)
Average glandular dose--The average absorbed dose accruing
to the glandular tissue of the breast.
(11)
Beam-limiting device--A device that provides a means to
restrict the dimensions of the x-ray field.
(12)
Breast implant--A prosthetic device implanted in the breast.
(13)
Calendar quarter--Any one of the following time periods
during a given year: January 1 - March 31, April 1 - June 30, July 1 - September
30, or October 1 - December 31.
(14)
Calibration of instruments--The comparative response or
reading of an instrument relative to a series of known radiation values over
the range of the instrument.
(15)
Category I continuing medical education units (CMEU)--Educational
activities designated as Category I and approved by the Accreditation Council
for Continuing Medical Education, the American Osteopathic Association, a
state medical society, or an equivalent organization.
(16)
Certification--An authorization for the use of a mammography
system or mammography machines used for interventional breast radiography.
(17)
Clinical image--See the definition for mammogram.
(18)
Contact hour--An hour of training received through direct
instruction.
(19)
Continuing education unit (CEU)--One contact hour of training.
(20)
Control panel--That part of the radiation machine control
upon which are mounted the switches, knobs, push buttons, and other hardware
necessary for setting the technique factors.
(21)
Direct instruction--Instruction that includes:
(A)
face-to-face interaction between instructor(s) and student(s),
as when the instructor provides a lecture, conducts demonstrations, or reviews
student performance; or
(B)
the administration and correction of student examinations
by an instructor(s) with subsequent feedback to the student(s).
(22)
Direct supervision--Oversight of operations that include
the following.
(A)
During joint interpretation of mammograms, the supervising
interpreting physician reviews, discusses, and confirms the diagnosis of the
physician being supervised and signs the resulting report before it is entered
into the patient's record.
(B)
During performance of a mammography examination, the supervising
medical radiologic technologist is present to observe and correct, as needed,
the individual who is performing the examination.
(C)
During performance of a survey of the registrant's equipment
and quality assurance program, the supervising medical physicist is present
to observe, and correct, as needed, the individual who is conducting the survey.
(23)
Established operating level--The value of a particular
quality assurance parameter that has been established as an acceptable normal
level by the registrant's quality assurance program.
(24)
Facility--A hospital, outpatient department, clinic, radiology
practice, mobile unit, an office of a physician, or other person that conducts
breast cancer screening or diagnosis through mammography activities, including
the following:
(A)
the operation of equipment to produce a mammogram;
(B)
processing of film;
(C)
initial interpretation of the mammogram; or
(D)
maintaining the viewing conditions for that interpretation.
(25)
FDA-approved accreditation body--An entity approved by
the FDA under Title 21, CFR, Part 900.3(d), to accredit mammography facilities.
(26)
Final assessment categories--The overall final assessment
of findings in a report of a mammography examination, classified in one of
the following categories:
(A)
"negative" indicates nothing to comment upon (if the interpreting
physician is aware of clinical findings or symptoms, despite the negative
assessment, these shall be explained);
(B)
"benign" is also a negative assessment;
(C)
"probably benign" indicates a finding(s) that has a high
probability of being benign;
(D)
"suspicious abnormality" indicates a finding(s) without
all the characteristic morphology of breast cancer but indicating a definite
probability of being malignant;
(E)
"highly suggestive of malignancy" indicates a finding(s)
that has a high probability of being malignant;
(F)
"known biopsy proven malignancy" indicates appropriate
action should be taken;
(G)
"post procedure mammogram" indicates a mammogram to confirm
the deployment and position of a breast tissue marker; or
(H)
"incomplete" indicates there is a need for additional imaging
evaluation and/or prior mammograms for comparison. Reasons why no assessment
can be made shall be stated by the interpreting physician.
(27)
First allowable time--The earliest time a resident physician
is eligible to take the diagnostic radiology boards from an FDA-designated
certifying body.
(28)
Formal training--Attendance and participation in direct
instruction. This does not include self-study programs.
(29)
Half-value layer (HVL)--The thickness of a specified material
that attenuates the beam of radiation to an extent such that the exposure
rate is reduced to one-half of its original value. In this definition, the
contribution of all scattered radiation, other than any that might be present
initially in the beam concerned, is deemed to be excluded.
(30)
Healing arts--Any system, treatment, operation, diagnosis,
prescription, or practice for the ascertainment, cure, relief, palliation,
adjustment, or correction of any human disease, ailment, deformity, injury,
or unhealthy or abnormal physical or mental condition.
(31)
Image receptor--Any device that transforms incident x-ray
photons either into a visible image or into another form that can be made
into a visible image by further transformations.
(32)
Institutional review board (IRB)--Any board, committee,
or other group formally designated by an institution to review, approve the
initiation of, and conduct periodic review of biomedical research involving
human subjects.
(33)
Interpreting physician--A licensed physician who interprets
mammographic images and who meets the requirements of subsection (r)(1) of
this section.
(34)
Interventional breast radiography--Imaging of a breast
during invasive interventions for localization or biopsy procedures.
(35)
Investigational device exemption--An exemption that allows
the investigational device to be used in a clinical study in order to collect
safety and effectiveness data required to support a Premarket Approval application
or a 501(k) Premarket Notification submission to FDA.
(36)
Kerma--The sum of the initial energies of all the charged
particles liberated by uncharged ionizing particles in a material of given
mass.
(37)
Laterality--The designation of either the right or left
breast.
(38)
Lead interpreting physician--The interpreting physician
assigned the general responsibility for ensuring that a facility's quality
assurance program meets all of the requirements of subsections (u), (v), and
(w) of this section.
(39)
Mammogram--A radiographic image produced through mammography.
(40)
Mammographic modality--A technology for radiography of
the breast. Examples are screen-film mammography and full-field digital mammography.
(41)
Mammography--The use of x-radiation to produce an image
of the breast that may be used to detect the presence of pathological conditions
of the breast. For the purposes of this section, mammography does not include
radiography of the breast performed as follows:
(A)
during invasive interventions for localization or biopsy
procedures except as specified in subsection (bb) of this section; or
(B)
with an investigational mammography device as part of a
scientific study conducted in accordance with FDA's investigational device
exemption regulations.
(42)
Mammography machine(s)--A unit consisting of components
assembled for the production of x-rays for use during mammography. These include,
at a minimum, the following:
(A)
an x-ray generator;
(B)
an x-ray control;
(C)
a tube housing assembly;
(D)
a beam limiting device; and
(E)
supporting structures.
(43)
Mammography medical outcomes audit--A systematic collection
of mammography results compared with outcomes data.
(44)
Mammography system--A system that includes the following:
(A)
an x-ray machine used as a source of radiation in producing
images of breast tissue;
(B)
an imaging system used for the formation of a latent image
of breast tissue;
(C)
an imaging-processing device for changing a latent image
of breast tissue to a visual image that can be used for diagnostic purposes;
(D)
a viewing device used for the visual evaluation of an image
of breast tissue if the image is produced in interpreting visual data captured
on an image receptor;
(E)
a medical radiologic technologist who performs mammography;
and
(F)
a physician who engages in mammography and who meets the
requirements of this section relating to the reading, evaluation, and interpretation
of mammograms.
(45)
Mandatory training--Additional training required by the
agency certifying body or FDA-approved accreditation body for interpreting
physicians, medical radiologic technologists, or medical physicists as the
result of a required corrective action.
(46)
Mean optical density--The average of the optical densities
measured using uniform, defect-free absorber thicknesses of 2, 4, and 6 centimeters
(cm) with values of kilovolt peak (kVp) clinically appropriate for those thicknesses.
(47)
Medical physicist--An individual who performs surveys
and evaluations of mammographic equipment and facility quality assurance programs
in accordance with this section and who meets the qualifications in subsection
(r)(3) of this section.
(48)
Medical radiologic technologist (operator of equipment)--An
individual specifically trained in the use of radiographic equipment and the
positioning of patients for radiographic examinations, who performs mammography
examinations in accordance with this section and who meets the qualifications
in subsection (r)(2) of this section.
(49)
Mobile service operation--The provision of mammography
machines and personnel at temporary sites for limited time periods.
(50) Multi-reading--Two or more physicians interpreting the
same mammogram. At least one physician shall be qualified as an interpreting
physician.
(51) Optical density (OD)--A measure of the fraction of incident
light transmitted through a developed film and defined by the equation:
Figure: 25 TAC §289.230(c)(51)
(52) Patient--Any individual who undergoes a mammography examination
in a facility, regardless of whether the person is referred by a physician
or is self-referred.
(53) Phantom--A test object used to simulate radiographic characteristics
of compressed breast tissue and containing components that radiographically
model aspects of breast disease and cancer.
(54) Phantom image--A radiographic image of a phantom.
(55) Physical science--This includes physics, chemistry, radiation
science (including medical physics and health physics), and engineering.
(56)
Positive mammogram--A mammogram that has an overall assessment
of findings that are either "suspicious" or "highly suggestive of malignancy."
(57)
Practitioner of the healing arts (practitioner)--For the
purposes of this section, a person licensed to practice healing arts by the
Texas Medical Board as a physician.
(58)
Provisional certification--A provisional authorization
described in subsection (g) of this section.
(59)
Qualified instructor--An individual whose training and
experience prepares him or her to carry out specified training assignments.
Interpreting physicians, medical radiologic technologists, or medical physicists
who meet the requirements of subsection (r) of this section would be considered
qualified instructors in their respective areas of mammography. Other examples
of individuals who may be qualified instructors for the purpose of providing
training to meet the requirements of this section include, but are not limited
to, instructors in a post-high school training institution and manufacturers'
representatives.
(60)
Quality control technologist--An individual meeting the
requirements of subsection (r)(2) of this section who is responsible for those
quality assurance responsibilities not assigned to the lead interpreting physician
or to the medical physicist.
(61)
Radiation machine--For the purposes of this part, radiation
machine also means mammography machine.
(62)
Self-referral mammography--The use of x-radiation to test
asymptomatic women for the detection of diseases of the breasts when such
tests are not specifically and individually ordered by a licensed physician.
(63)
Serious adverse event--An adverse event that may significantly
compromise clinical outcomes, or an adverse event for which a facility fails
to take appropriate corrective action in a timely manner.
(64)
Serious complaint--A report of a serious adverse event.
(65)
Source-to-image receptor distance (SID)--The distance
from the source to the center of the input surface of the image receptor.
(66)
Standard breast--A 4.2 cm thick compressed breast consisting
of 50% glandular and 50% adipose tissue.
(67)
Survey--An on-site physics consultation and evaluation
of a facility quality assurance program performed by a medical physicist.
(68)
Technique chart--A chart that provides all necessary generator
control settings and geometry needed to make clinical radiographs.
(69)
Traceable to a national standard--Calibrated at either
the National Institute of Standards and Technology (NIST) or at a calibration
laboratory that participates in a proficiency program with NIST at least once
every two years. The results of the proficiency test conducted within 24 months
of calibration shall show agreement within plus or minus 3.0% of the national
standard in the mammography energy range.
(d)
Prohibitions.
(1)
Radiographic equipment designed for general purpose or
special nonmammography procedures shall not be used for mammography. This
includes systems that have been modified or equipped with special attachments
for mammography.
(2)
The agency may prohibit use of mammography machines that
pose a significant threat or endanger public health and safety, in accordance
with §289.231 of this title and §289.205 of this title.
(3)
Individuals shall not be exposed to the useful beam except
for healing arts purposes and unless such exposure has been authorized by
a licensed physician. This provision specifically prohibits intentional exposure
for the following purposes:
(A)
exposure of an individual for training, demonstration,
or other non-healing arts purposes;
(B)
exposure of an individual for the purpose of healing arts
screening (self referral mammography) except as authorized by subsection (cc)
of this section; and
(C)
exposure of an individual for the purpose of research except
as authorized by subsection (dd) of this section.
(e)
Exemptions.
(1)
Mammography machines or cabinet x-ray machines used exclusively
for examination of breast biopsy specimens are exempt from the requirements
of this section. These machines are required to meet applicable provisions
of §289.226 of this title and §289.228 of this title (relating to
Radiation Safety Requirements for Analytical and Other Industrial Radiation
Machines).
(2)
Mammography machines used exclusively for interventional
breast radiography are exempt from the requirements of this section except
for those listed in subsection (bb) of this section. These machines are not
required to be accredited by an FDA-approved accreditation body.
(3)
Loaner machines as described in subsection (n)(5) of this
section are exempt from the inspection requirements in subsection (gg) of
this section. These machines are not required to be accredited by an FDA-approved
accreditation body.
(4)
Mammography machines with investigational device exemptions
as described in subsection (dd) of this section and used in clinical studies
are exempt from the requirements of this chapter. These machines are not required
to be accredited by an FDA-approved accreditation body.
(5)
All mammography registrants are exempt from the posting
of radiation area requirements of §289.231(x) of this title provided
that the operator has continuous surveillance and access control of the radiation
area.
(f)
Requirements for mammography systems certification.
(1)
To obtain a certification, facilities shall meet the quality
standards in subsections (r) - (aa) of this section and be accredited by an
FDA-approved accreditation body. In order to qualify for certification, new
facilities must apply to the agency certifying body in accordance with the
following requirements and to an FDA-approved accreditation body and receive
acceptance of the accreditation application. If the facility chooses to be
accredited by the agency accreditation body, the facility shall submit the
information required in this subsection and §289.234(d) of this title.
(2)
Each person having a mammography machine shall submit an
application in accordance with §289.226(e)(1) - (3) and (5) - (7) and
(f)(4) - (5) of this title, and receive certification from the agency certifying
body before beginning use of the mammography machine on humans.
(3)
An application for certification shall be signed by a licensed
physician. The signature of the applicant and the radiation safety officer
(RSO) shall also be required.
(4)
An application for certification may contain information
on multiple mammography machines. Each mammography machine must be identified
by referring to the machine's manufacturer, model number, and serial number
of the control panel. If this is not a new certification, the registrant shall
maintain and provide proof of current accreditation. If accreditation expires
before the expiration of the certification, the registrant shall submit proof
of renewed status to the agency.
(5)
Each applicant shall submit documentation of the following:
(A)
personnel qualifications, including dates of licensure
or certification, in accordance with subsection (r) of this section;
(B)
manufacturer, model, and serial number of each mammography
machine control panel;
(C)
evidence that a medical physicist:
(i)
has determined that each machine meets the equipment standards
in subsection (s) of this section;
(ii)
has performed a survey and a mammography equipment evaluation
in accordance with subsection (v)(10) and (11) of this section; and
(iii)
has determined that the average glandular dose for one
craniocaudal-caudal view for each machine does not exceed the value in subsection
(v)(5)(F) of this section;
(D)
self-referral program information in accordance with subsection
(cc) of this section, if the facility offers self-referral mammography; and
(E)
items required for authorization of a mobile service operation
in accordance with §289.226(g) of this title, if the facility provides
a mobile service.
(g)
Issuance of certification and provisional certification.
(1)
Certification. A certification will be issued if the agency
certifying body determines that an application meets the requirements of the
Act and the requirements of this chapter. The certification authorizes the
proposed activity in such form and contains such conditions and limitations
as the agency certifying body deems appropriate or necessary. The certification
may include one or both of the following:
(A)
mammography systems and facilities certification, following
approval of accreditation by an FDA-approved accreditation body; or
(B)
certification of mammography machines used for interventional
breast radiography.
(2)
Requirements and conditions. The agency certifying body
may incorporate in the certification at the time of issuance, or thereafter
by amendment, such additional requirements and conditions with respect to
the registrant's possession, use, and transfer of radiation machines subject
to this chapter as it deems appropriate or necessary in order to:
(A)
minimize danger to occupational and public health and safety;
(B)
require additional reports and the keeping of additional
records as may be appropriate or necessary; and
(C)
prevent loss or theft of radiation machines subject to
this section.
(3)
Additional information. The agency certifying body may
request, and the registrant shall provide, additional information after the
certification has been issued to enable the agency certifying body to determine
whether the certification should be modified in accordance with §289.226(r)
of this title.
(4)
Provisional certification application. A new facility is
eligible to apply for a provisional certification. The provisional certification
will enable the facility to perform mammography and to obtain the clinical
images needed to complete the accreditation process. To apply for and receive
a provisional certification, a facility must meet the requirements of this
chapter and submit the necessary information to an FDA-approved accreditation
body. If the facility chooses to be accredited by the agency accreditation
body, the facility shall submit the information required in subsection (f)
of this section and §289.234(d) of this title to the agency accreditation
body.
(5)
Issuing provisional certifications. Following the agency
certifying body's receipt of the accreditation body's decision that a facility
has submitted the required information, the agency certifying body may issue
a provisional certification to a facility upon determination that the facility
has satisfied the requirements of the Act and this chapter. A provisional
certification shall be effective for up to six months from the date of issuance.
A provisional certification cannot be renewed, but a facility may apply for
a 90-day extension of the provisional certification.
(6)
Extension of provisional certification. Extension of provisional
certifications shall be in accordance with the following.
(A)
To apply for a 90-day extension to a provisional certification,
a facility shall submit to the FDA-approved accreditation body who issued
the original certificate, a statement of what the facility is doing to obtain
certification and evidence that there would be a significant adverse impact
on access to mammography in the geographic area served if such facility did
not obtain an extension.
(B)
The agency certifying body may issue a 90-day extension
for a provisional certification upon determination that the extension meets
the criteria in paragraph (4) of this subsection.
(C)
There can be no renewal of a provisional certification
beyond the 90-day extension.
(7)
Reinstatement policy. A previously certified facility that
has allowed its certification to expire, that has been refused a renewal of
its certification by the agency certifying body, or that has had its certification
suspended or revoked by the agency certifying body, may reapply to have the
certification reinstated so that the facility may be considered to be a new
facility and thereby be eligible for a provisional certification.
(A)
Unless prohibited from reinstatement under subsection (h)(5)
of this section, a facility applying for reinstatement shall:
(i)
contact an FDA-approved accreditation body for reapplication
for accreditation;
(ii)
fully document its history as a previously provisionally
certified or certified mammography facility, including the following information:
(I)
name and address of the facility under which it was previously
provisionally certified or certified;
(II)
name of previous owner/lessor;
(III)
facility identification number assigned to the facility
under its previous certification by the FDA or the agency certifying body;
and
(IV)
expiration date of the most recent FDA or agency provisional
certification; and
(iii)
justify application for reinstatement of accreditation
by submitting to an FDA-approved accreditation body a corrective action plan
that details how the facility has corrected deficiencies that contributed
to the lapse of, denial of renewal, or revocation of its certification.
(B)
The agency certifying body may issue a provisional certification
to the facility if the agency determines that the facility:
(i)
has adequately corrected, or is in the process of correcting,
pertinent deficiencies; and
(ii)
has taken sufficient corrective action since the lapse
of, denial of renewal, or revocation of its previous certification.
(C)
After receiving the provisional certification, the facility
may lawfully perform mammography while completing the requirements for accreditation
and certification.
(h)
Suspension or revocation of certification.
(1)
Except as provided in paragraph (2) of this subsection,
the agency certifying body may suspend or revoke a certification issued by
the agency certification body if it finds, after providing the owner or operator
of the facility with notice and opportunity for a hearing in accordance with §289.205
of this title, that the owner, operator, or any employee of the facility:
(A)
has been guilty of misrepresentation in obtaining the certification;
(B)
has failed to comply with the requirements of this chapter;
(C)
has failed to comply with requests of the agency certifying
body or an FDA-approved accreditation body for records, information, reports,
or materials that are necessary to determine the continued eligibility of
the facility for a certification or continued compliance with the requirements
of this chapter;
(D)
has refused a request of a duly designated FDA inspector,
state inspector, or an FDA-approved accreditation body representative for
permission to inspect the facility or the operations and pertinent records
of the facility;
(E)
has violated or aided and abetted in the violation of any
provision of or regulation promulgated pursuant to the requirements of the
Act and the requirements of this chapter; or
(F)
has failed to comply with prior sanctions imposed by the
agency certifying body under §289.205 of this title.
(2)
The agency certifying body may suspend a certification
of a facility before holding a hearing if it makes a finding described in
paragraph (1) of this subsection and also determines that:
(A)
the failure to comply with requirements presents a serious
risk to human health;
(B)
the refusal to permit inspection makes immediate suspension
necessary; or
(C)
there is reason to believe that the violation or aiding
and abetting of the violation was intentional or associated with fraud.
(3)
If the agency certifying body suspends a certification
in accordance with paragraph (2) of this subsection:
(A)
the agency certifying body shall provide the facility with
an opportunity for a hearing under §289.205 not later than 60 days from
the effective date of this suspension; and
(B)
the suspension shall remain in effect until the agency
certifying body determines that:
(i)
allegations of violations or misconduct were not substantiated;
(ii)
violations of requirements have been corrected to the
agency certifying body's satisfaction; or
(iii)
the certification is revoked in accordance with paragraph
(4) of this subsection.
(4)
After providing a hearing in accordance with paragraph
(3)(A) of this subsection, the agency certifying body may revoke the certification
if the agency determines that the facility:
(A)
is unwilling or unable to correct violations that were
the basis for suspension; or
(B)
has engaged in fraudulent activity to obtain or continue
certification.
(5)
If a facility's certification was revoked on the basis
of an act described in §289.205 of this title, no person who owned or
operated that facility at the time the act occurred may own or operate a mammography
facility within two years of the date of revocation.
(i)
Appeal of adverse accreditation or reaccreditation decisions
that preclude certification or recertification.
(1)
The appeal process described in this subsection is available
only for adverse accreditation or reaccreditation decisions that preclude
certification by the agency certifying body. Agency certifying body decisions
to suspend or revoke certificates that are already in effect will be handled
in accordance with subsection (h) of this section.
(2)
Upon learning that a facility has failed to become accredited
or reaccredited, the agency certifying body will notify the facility that
the agency certifying body is unable to certify that facility without proof
of accreditation.
(3)
A facility that has been denied accreditation or reaccreditation
and cannot achieve satisfactory resolution of an adverse accreditation decision
through the FDA-approved accreditation body's appeal process is entitled to
further appeal to the FDA.
(4)
A facility cannot perform mammography services while an
adverse accreditation decision is being appealed.
(j)
Denial of certification.
(1)
The agency certifying body may deny the application if
the agency certifying body has reason to believe that:
(A)
the facility will not be operated in accordance with the
provisions of subsections (r) - (aa) of this section;
(B)
the facility will not permit inspections or provide access
to records or information in a timely fashion;
(C)
any material false statement in the application or any
statement of fact required under provision of the Act was made;
(D)
conditions revealed by such application or statement of
fact or any report, record, inspection, or other means that would warrant
the agency certification body to refuse to grant a certification of mammography
facility on an original application; or
(E)
the facility failed to observe any of the terms and conditions
of the Act, this chapter, or order of the agency.
(2)
Before the agency certification body denies an application
for certification, the agency shall give notice of the denial, the facts warranting
the denial, and shall afford the applicant an opportunity for a hearing in
accordance with §289.205(h) of this title. If no request for a hearing
is received by the director of the Radiation Control Program within 30 days
of date of receipt of the notice, the agency may proceed to deny. The applicant
shall have the burden of proof showing cause why the application should not
be denied.
(3)
If the agency certifying body denies an application for
certification by a facility that has received accreditation from an FDA-approved
accreditation body, the agency certifying body shall provide the facility
with a written statement of the grounds on which the denial is based.
(k)
Appeals of denial of certification.
(1)
The appeals procedures described in this subsection are
available only to facilities that are denied certification by the agency certifying
body after they have been accredited by an FDA-approved accreditation body.
Appeals for facilities that have failed to become accredited with the agency
accreditation body shall be in accordance with §289.234(h) of this title.
(2)
A facility that has been denied certification may request
reconsideration and appeal of the agency certifying body's determination in
accordance with the applicable provisions of §289.205 of this title.
(l)
Modification of certification. Modification of certification
shall be in accordance with §289.226(r) of this title.
(m)
Specific terms and conditions of certification. Specific
terms and conditions of certification shall be in accordance with §289.226(l)
of this title.
(n)
Responsibilities of registrant.
(1)
In addition to the requirements of §289.226(m)(3)
- (7) of this title, a registrant shall notify the agency certifying body
in writing prior to any changes that would render the information contained
in the application or the certification inaccurate. These include but are
not limited to the following:
(A)
name and mailing address;
(B)
street address where machine(s) will be used; and
(C)
mammography machines.
(2)
Prior to employing the individuals listed in subparagraphs
(A) - (E) of this paragraph, the registrant is required to verify and maintain
copies of their qualifications. Documentation of qualifications of individuals
listed in subparagraphs (A) - (E) of this paragraph and notification of a
change in any of the following is required within 30 days of such change:
(A)
radiation safety officer;
(B)
lead interpreting physician;
(C)
interpreting physicians;
(D)
medical radiologic technologists; or
(E)
medical physicist.
(3)
Registrants utilizing interpreting physicians or technologists
from a temporary service shall verify and maintain copies of the qualifications
of these individuals for inspection by the agency. The registrant does not
need to notify the agency certifying body unless these personnel will be at
the facility for a period exceeding four weeks.
(4)
All mammography facilities installing new or replacement
mammography machines shall have either current accreditation or have submitted
an application to an FDA-approved accreditation body for review unless exempted
by subsection (e)(1) - (3) of this section. A mammography machine shall not
be used to perform mammograms if an application for accreditation for that
machine has been denied, or if the accreditation has been suspended or expired.
(5)
A facility with an existing certification may begin using
a new or replacement machine before receiving an updated certification if
the registrant submits to the agency certifying body and to the FDA-approved
accreditation body, documentation with a medical physicist's report in accordance
with subsection (v)(10) and (11) of this section, verifying compliance of
the new machine with this section. The medical physicist's report is required
prior to using the machine on patients.
(6)
Loaner mammography machines may be used on patients for
60 days without adding the mammography machine to the certification. A medical
physicist's report verifying compliance of the loaner mammography machine
with this section shall be completed prior to use on patients. The results
of the survey must be submitted to the agency with a cover letter indicating
period of use. If the use period will exceed 60 days, the facility shall add
the mammography machine to its certification and a fee will be assessed.
(7)
Records of training and experience and all other records
required by this section shall be maintained for review in accordance with
subsection (ff) of this section.
(o)
Renewal of certification.
(1)
A certification for a mammography system is valid for three
years from the date of issuance unless the certification of the facility is
suspended or revoked prior to such deadlines.
(2)
A mammography facility filing an application for renewal
in accordance with subsection (f) of this section and fees in accordance with §289.204
of this title before the existing certification expires:
(A)
may continue to perform mammography until the expiration
date of the certification; or
(B)
if the facility receives written authorization from an
FDA-approved accreditation body, may continue to perform mammography until
the review process is complete and the accreditation status has been determined
by the FDA-approved accreditation body.
(3)
A facility with mammography machines used for interventional
breast radiography shall file an application for renewal in accordance with
subsection (bb)(9) of this section and pay the fee required by §289.204
of this title.
(p)
Expiration of certification.
(1)
Except as provided by subsection (o) of this section, each
certification expires at the end of the day in the month and year stated on
the certificate of registration. Expiration of the certification does not
relieve the registrant of the requirements of this chapter.
(2)
If a registrant does not submit an application for renewal
of the certification under subsection (o) of this section, as applicable,
the registrant shall on or before the expiration date specified in the certification:
(A)
terminate use of all mammography machines;
(B)
notify the agency certifying body in writing of the film
storage location of mammography patients' films and address how the requirements
of subsection (t)(4) of this section will be met;
(C)
pay any outstanding fees in accordance with §289.204
of this title; and
(D)
submit a record of the disposition of the mammography machine(s)
to the agency certifying body. If the machine(s) was transferred, include
to whom it was transferred.
(q)
Termination of certification. When a registrant decides
to terminate all activities involving mammography machines authorized under
the certification, the registrant shall:
(1)
notify the agency certifying body and the FDA-approved
accreditation body immediately;
(2)
request termination of the certification in writing;
(3)
pay any outstanding fees in accordance with §289.204
of this title;
(4)
notify the agency certifying body, in writing, of the film
storage location of mammography patients' films and address how the requirements
of subsection (t)(4) of this section will be met; and
(5)
submit a record of the disposition of the mammography machine(s)
to the agency certifying body. If the machine(s) was transferred, include
to whom it was transferred.
(r)
Personnel qualifications. The following requirements apply
to all personnel involved in any aspect of mammography, including the production
and interpretation of mammograms.
(1)
Interpreting physician. Each physician interpreting mammograms
shall hold a current Texas license issued by the Texas Medical Board and meet
the following qualifications.
(A)
Initial qualifications. Before interpreting mammograms
independently, the physician shall:
(i)
be certified by the American Board of Radiology, the American
Osteopathic Board of Radiology, or one of the other bodies approved by the
FDA to certify interpreting physicians or have at least three months of documented
formal training in the interpretation of mammograms and in topics related
to mammography in accordance with subsection (hh)(2) of this section;
(ii)
have had a minimum of 60 hours of documented category
I CMEUs in mammography. At least 15 of the 60 hours shall have been acquired
within three years immediately prior to the date that the physician qualified
as an interpreting physician. Hours spent in residency specifically devoted
to mammography will be equivalent to category I CMEUs and accepted if documented
in writing by the appropriate representative of the training institution;
and
(iii)
have interpreted or multi-read, under the direct supervision
of an interpreting physician, at least 240 mammographic examinations within
the six-month period immediately prior to the date that the physician qualifies
as an interpreting physician.
(B)
Exemptions.
(i)
Physicians who qualified as interpreting physicians in
accordance with the requirements of §289.230 that were in effect prior
to April 28, 1999, or any other equivalent state or federal requirements in
effect prior to April 28, 1999, are considered to have met the initial requirements
of subparagraph (A) of this paragraph.
(ii)
Physicians who have interpreted or multi-read at least
240 mammographic examinations under the direct supervision of an interpreting
physician in any six month period during the last two years of a diagnostic
radiology residency and who became board certified at the first allowable
time, are exempt from subparagraph (A)(iii) of this paragraph.
(C)
Continuing education and experience. The time period for
completing continuing education is a 36-month period and the time period for
completing continuing experience is a 24-month period. These periods begin
when a physician completes the requirements to become an interpreting physician
in subparagraph (A) of this paragraph. The facility shall choose one of the
dates in clause (i) of this subparagraph to determine the 36-month continuing
education period and one of the dates in clause (ii) of this subparagraph
to determine the 24-month continuing experience period. Each interpreting
physician shall maintain qualifications by meeting the following requirements:
(i)
participating in education programs by completing at least
15 category I CMEUs in mammography that shall include at least six CMEUs in
each modality used by the interpreting physician in his/her practice or by
teaching mammography courses. CMEUs earned through teaching a specific course
can be counted only once during the 36-month period. The continuing education
must be completed in the 36 months immediately preceding:
(I)
the date of the registrant's annual inspection;
(II)
the last day of the calendar quarter preceding the inspection;
or
(III)
any date in between the two;
(ii)
interpreting or multi-reading at least 960 mammographic
examinations that must be completed during the 24 months immediately preceding:
(I)
the date of the registrant's annual inspection;
(II)
the last day of the calendar quarter preceding the inspection;
or
(III)
any date in between the two; and
(iii)
accumulating at least eight hours of CMEUs in any mammography
modality in which the interpreting physician has not been previously trained,
prior to independently using the new modality.
(D)
Re-establishing qualifications. Before resuming independent
interpretation of mammograms, interpreting physicians who fail to maintain
the required continuing education or experience requirements shall re-establish
their qualifications by completing one or both of the following requirements,
as applicable:
(i)
obtain a sufficient number of additional category I CMEUs
to bring their total up to the 15 category I CMEU credits required in the
previous 36 months; and
(ii)
within the six months immediately prior to resuming independent
interpretation and under the direct supervision of an interpreting physician,
interpret or multi-read one of the following, whichever is less:
(I)
at least 240 mammographic examinations; or
(II)
a sufficient number of mammographic examinations to bring
the total up to 960 examinations for the prior 24 months.
(E)
Any mandatory training required by the agency certifying
body or an FDA-approved accreditation body shall be completed prior to independently
interpreting mammograms. Records of any mandatory training shall be maintained
in accordance with subsection (ff)(3) of this section.
(2)
Medical radiologic technologists (operators of equipment).
Each person performing mammographic examinations shall have current certification
as a medical radiologic technologist under the Medical Radiologic Technologist
Certification Act, Texas Occupations Code, Chapter 601, and shall meet the
following qualifications.
(A)
Initial requirements. Before performing mammographic examinations,
the operator of equipment shall have:
(i)
completed a minimum of 40 contact hours of training as
outlined in subsection (hh)(1) of this section by a qualified instructor;
and
(ii)
performed a minimum of 25 mammographic examinations under
the direct supervision of an individual qualified in accordance with the requirements
of this paragraph. The 25 mammographic examinations may be obtained concurrently
with the 40 contact hours of training specified in clause (i) of this subparagraph
but shall not exceed 16 hours of the 40 contact hours.
(B)
Exemptions. Equipment operators who qualified as medical
radiologic technologists to perform mammography in accordance with the requirements
of §289.230 that were in effect prior to April 28, 1999, and any other
federal requirements in effect prior to April 28, 1999, are considered to
have met the initial requirements of subparagraph (A) of this paragraph.
(C)
Continuing education and experience. The time period for
completing continuing education is a 36-month period and the time period for
completing continuing experience is a 24-month period. The period for continuing
education begins when a technologist completes the requirements in subparagraph
(A) of this paragraph. The period for continuing experience begins when a
technologist completes the requirements in subparagraph (A) of this paragraph,
or April 28, 1999, whichever is later. The facility shall choose one of the
dates in clause (i) of this subparagraph to determine the 36-month continuing
education period and one of the dates in clause (ii) of this subparagraph
to determine the 24-month continuing experience period. Each medical radiologic
technologist shall maintain qualifications by meeting the following requirements:
(i)
participating in education programs by completing at least
15 CEUs in mammography that shall include at least six CEUs in each modality
used by the technologist or by teaching mammography courses. CEUs earned through
teaching a specific course can be counted only once during the 36-month period.
The continuing education must be completed in the 36 months immediately preceding:
(I)
the date of the registrant's annual inspection;
(II)
the last day of the calendar quarter preceding the inspection;
or
(III)
any date in between the two;
(ii)
performing a minimum of 200 mammographic examinations
that must be completed during the 24 months immediately preceding:
(I)
the facility's annual inspection;
(II)
the last day of the calendar quarter preceding the inspection;
or
(III)
any date in between the two; and
(iii)
accumulating at least eight hours of CEUs in any mammography
modality in which the medical radiologic technologist has not been previously
trained, prior to independently using the new modality.
(D)
Requalification. Before resuming independent performance
of mammograms, medical radiologic technologists who fail to maintain the continuing
education or experience requirements shall re-establish their qualifications
by completing one or both of the following requirements, as applicable:
(i)
obtaining a sufficient number of additional CEUs to bring
their total up to the 15 CEU credits required in the previous 36 months, at
least six of which shall be related to each modality used by the technologist
in mammography; and/or
(ii)
performing a minimum of 25 mammographic examinations under
the direct supervision of a qualified medical radiologic technologist.
(E)
Any mandatory training required by the agency certifying
body or an FDA-approved accreditation body shall be completed prior to independently
performing mammograms. Records of any mandatory training shall be maintained
in accordance with subsection (ff)(3) of this section.
(3)
Medical physicist. Each medical physicist performing mammographic
surveys, evaluating mammographic equipment, or providing oversight of the
facility quality assurance program in accordance with subsection (u) of this
section, shall hold a current Texas license under the Medical Physics Practice
Act, Texas Occupations Code, Chapter 602, in diagnostic radiological physics
and be registered with the agency or employed by an entity registered with
the agency, in accordance with §289.226(j) of this title and the Act,
unless exempted by §289.226(d)(6) of this title. Each medical physicist
shall meet the following qualifications.
(A)
Initial qualifications. Before performing surveys and evaluating
mammographic equipment independently, the medical physicist shall:
(i)
have a masters degree or higher in a physical science from
an accredited institution, with no less than 20 semester hours or equivalent
(30 quarter hours) of college undergraduate or graduate level physics;
(ii)
have 20 contact hours of documented specialized training
in conducting surveys of mammography facilities; and
(iii)
have experience conducting surveys of at least one mammography
facility and a total of at least ten mammography machines. After April 28,
1999, experience conducting surveys must be acquired under the direct supervision
of a medical physicist who meets the requirements of subparagraphs (A) and
(C) of this paragraph. No more than one survey of a specific machine within
a period of 60 days can be counted towards the total mammography machine survey
requirement.
(B)
Alternative initial qualifications. Individuals who qualified
as a medical physicist in accordance with the requirements of this section
that were in effect prior to April 28, 1999, or any other equivalent state
or federal requirements in effect prior to April 28, 1999, and have met the
following additional qualifications prior to April 28, 1999, are determined
to have met the initial qualifications of subparagraph (A) of this paragraph:
(i)
a bachelor's degree or higher in a physical science from
an accredited institution with no less than ten semester hours or equivalent
of college undergraduate or graduate level physics;
(ii)
40 contact hours of documented specialized training in
conducting surveys of mammography facilities; and
(iii)
experience conducting surveys of at least one mammography
facility and a total of at least 20 mammography machines. No more than one
survey of a specific machine within a period of 60 days can be counted towards
the total mammography machine survey requirement. The training and experience
requirements must be met after fulfilling the degree requirements.
(C)
Continuing education and experience. The time period for
completing continuing education is a 36-month period and the time period for
completing continuing experience is a 24-month period. The period for continuing
education will begin when a physicist completes the requirements in subparagraph
(A) of this paragraph. The time period for continuing experience will begin
when a physicist completes the requirements in subparagraph (A) of this paragraph,
or April 28, 1999, whichever is later. The facility shall choose one of the
dates in clause (i) of this subparagraph to determine the 36-month continuing
education period and one of the dates in clause (ii) of this subparagraph
to determine the 24-month continuing experience period. Each medical physicist
shall maintain his/her qualifications by meeting the following requirements:
(i)
participating in education programs, either by teaching
or completing at least 15 CEUs in mammography that shall include hours of
training appropriate to each mammographic modality evaluated by the medical
physicist during his or her surveys. CEUs earned through teaching a specific
course can be counted only once during the 36-month period. The continuing
education must be completed in the 36 months immediately preceding:
(I)
the date of the registrant's annual inspection;
(II)
by the last day of the calendar quarter preceding the
inspection; or
(III)
any date in between the two;
(ii)
performing surveys of two mammography facilities and a
total of at least six mammography machines (no more than one survey of a specific
facility within a ten-month period or a specific machine within a period of
60 days can be counted towards the total mammography machine survey requirement).
The continuing experience must be completed during the 24 months immediately
preceding:
(I)
the date of the facility's annual inspection;
(II)
by the last day of the calendar quarter preceding the
inspection; or
(III)
any date in between the two; and
(iii)
accumulating at least eight hours of CEUs in any mammography
modality in which the medical physicist has not been previously trained, prior
to independently using the new modality.
(D)
Re-establishing qualifications. Before resuming independent
performance of surveys and equipment evaluations, medical physicists who fail
to maintain the continuing education or experience requirements shall reestablish
their qualifications by completing one or both of the following requirements,
as applicable:
(i)
obtaining a sufficient number of additional CEUs to bring
their total up to the 15 CEU credits required in the previous 36 months; and
(ii)
performing a sufficient number of surveys, under the direct
supervision of a qualified medical physicist, to bring their total up to two
mammography facilities and a total of at least six mammography machines for
the prior 24 months. No more than one survey of a specific machine within
a period of 60 days shall be counted towards the total mammography machine
survey requirement.
(E)
Any mandatory training required by the agency certifying
body or an FDA-approved accreditation body shall be completed prior to independently
performing mammographic surveys, evaluating mammographic equipment, or providing
oversight of a facility's quality assurance program. Records of any mandatory
training shall be maintained in accordance with subsection (ff)(3) of this
section.
(4)
Retention of personnel records. Records documenting the
qualifications, continuing education, and experience of personnel in subsection
(r)(1) - (3) of this section shall be maintained for inspection by the agency
in accordance with subsection (ff) of this section.
(s)
Equipment standards. Only systems meeting the following
standards shall be used.
(1)
System design. The equipment shall have been specifically
designed and manufactured for mammography and in accordance with Title 21,
CFR, §§1010.2, 1020.30, and 1020.31.
(2)
Motion of tube-image receptor assembly. The assembly shall
be capable of being fixed in any position where it is designed to operate.
Once fixed in any such position, it shall not undergo unintended motion. In
the event of power interruption, this mechanism shall not fail.
(3)
Image receptors. Systems using screen-film image receptors
shall, at a minimum, provide for the following:
(A)
operation with image receptors of 18 x 24 cm and 24 x 30
cm;
(B)
operable moving grids matched to all image receptor sizes
provided;
(C)
operation with the grid removed from between the source
and image receptor for systems used for magnification procedures; and
(D)
image receptors to rest, post-loading, 15 minutes between
exposures.
(4)
Magnification. Systems used to perform noninterventional
problem solving procedures shall have radiographic magnification capability
available for use with, at a minimum, at least one magnification value within
the range of 1.4 to 2.0.
(5)
Focal spot and target material selection. Selection of
the focal spot or target material shall be as follows.
(A)
When more than one focal spot is provided, the system shall
indicate, prior to exposure, which focal spot is selected.
(B)
When more than one target material is provided, the system
shall indicate, prior to exposure, the preselected target material.
(C)
When the target material and/or focal spot is selected
by a system algorithm that is based on the exposure or on a test exposure,
the system shall display, after the exposure, the target material and/or focal
spot actually used during the exposure.
(6)
Compression. All mammography systems shall incorporate
a compression device.
(A)
Application of compression. Effective October 28, 2002,
and thereafter, each system shall provide the following features operable
from both sides of the patient:
(i)
an initial power-driven compression activated by hands-free
controls; and
(ii)
fine adjustment compression controls.
(B)
Compression paddle.
(i)
Systems shall be equipped with different sized compression
paddles that match the sizes of all full-field image receptors provided for
the system.
(ii)
Compression paddles for special purposes, including those
smaller than the full size of the image receptor (for example, spot compression)
may be provided. Such paddles are not subject to the requirements of clauses
(v) and (vi) of this subparagraph.
(iii)
Except as provided in clause (iv) of this subparagraph,
the compression paddle shall be flat and parallel to the breast support table
and shall not deflect from parallel by more than 1.0 cm at any point on the
surface of the compression paddle when compression is applied.
(iv)
Equipment intended by the manufacturer's design to not
be flat and parallel to the breast support table during compression shall
meet the manufacturer's design specifications and maintenance requirements.
(v)
The chest wall edge of the compression paddle shall be
straight and parallel to the edge of the image receptor.
(vi)
The chest wall edge may be bent upward to allow for patient
comfort, but shall not appear on the image.
(7)
Technique factor selection and display. Technique factor
selection and display shall be as follows.
(A)
Manual selection of milliampere seconds (mAs) or at least
one of its component parts, milliampere (mA) and/or time, shall be available.
(B)
The technique factors (peak tube potential in kilovolts
(kV) and either tube current in mA and exposure time in seconds or the product
of tube current and exposure time in mAs) to be used during an exposure shall
be indicated before the exposure begins, except when automatic exposure control
(AEC) is used, in which case the technique factors that are set prior to the
exposure shall be indicated.
(C)
When the AEC mode is used, the system shall indicate the
actual kVp and mAs used during the exposure. The mAs may be displayed as mA
and time.
(8)
Automatic exposure control. Each screen-film system shall
provide an AEC mode that is operable in all combinations of equipment configuration
provided, for example, contact, magnification, and various image receptor
sizes.
(A)
The positioning or selection of the detector shall permit
flexibility in the placement of the detector under the target tissue.
(i)
The size and available positions of the detector shall
be clearly indicated at the x-ray input surface of the breast compression
paddle.
(ii)
The selected position of the detector shall be clearly
indicated.
(B)
The system shall provide means to vary the selected optical
density from the normal (zero) setting.
(9)
X-ray film. The registrant shall use x-ray film for mammography
that has been designated by the film manufacturer as appropriate for mammography.
(10)
Intensifying screens. The registrant shall use intensifying
screens for mammography that have been designated by the screen manufacturer
as appropriate for mammography and shall use film that is matched to the screen's
spectral output as specified by the manufacturer.
(11)
Film processing solutions. For processing mammography
films, the registrant shall use chemical solutions that are capable of developing
the films used by the facility in a manner equivalent to the minimum requirements
specified by the film manufacturer.
(12)
Lighting. The registrant shall make available special
lights for film illumination (hot lights) capable of producing light levels
greater than that provided by the view box.
(13)
Film masking devices. Registrants shall ensure that film
masking devices that can limit the illuminated area to a region equal to or
smaller than the exposed portion of the film are available to all interpreting
physicians interpreting for the facility.
(14)
Equipment variances. Registrants with mammography equipment
that has been issued variances by FDA to Title 21, CFR, §§1020.2,
1020.30, 1020.31, or has had an alternative for a quality standard for equipment
approved by the FDA under the provisions of Title 21, CFR, §900.18, shall
maintain copies of those variances or alternative standards.
(15)
Light fields. For any mammography system with a light
beam that passes through the x-ray beam-limiting device, the light shall provide
an average illumination of not less than 160 lux (15 foot candles) at 100
cm or the maximum source-image receptor distance (SID), whichever is less.
(t)
Medical records and mammography reports.
(1)
Contents and terminology. Each registrant shall prepare
a written report of the results of each mammography examination that shall
include the following information:
(A)
name of the patient and an additional patient identifier;
(B)
date of the examination;
(C)
name and signature of the interpreting physician who interpreted
the mammogram (electronic signatures are acceptable);
(D)
overall final assessment of findings using the final assessment
categories as defined in subsection (c) of this section; and
(E)
recommendations made to the physician about what additional
actions, if any, should be taken. All clinical questions raised by the referring
physician shall be addressed in the report to the extent possible, even if
the assessment is negative or benign.
(2)
Communication of mammography results to the patient and
health care providers or physicians, as applicable. Each registrant shall
send reports as soon as possible, but no later than 30 days from the date
of the mammography examination, to:
(A)
patients advising them of the results of the mammography
examination and any further medical needs indicated. The report shall include
a summary written in language easily understood by a lay person; and
(B)
referring physicians, or in the case of self-referral,
to the physician indicated by the patient, advising them of the results of
the mammography examination, containing the information specified in paragraph
(1) of this subsection, and any further medical needs indicated.
(3)
Follow-up with patients and physicians. Each registrant
shall follow-up to confirm the following:
(A)
that patients with positive findings and patients needing
repeat exams have received proper notification; and
(B)
that physicians have received proper notification of patients
with positive findings or needing repeat exams.
(4)
Retention of clinical images.
(A)
Each registrant that performs mammograms shall maintain
mammography films and reports in a permanent medical record for a minimum
of five years. If no additional mammograms of the patient are performed at
the facility, the films and reports shall be maintained for a minimum of ten
years.
(B)
Each registrant that performs mammograms shall, within
30 days of request by or on behalf of the patient, permanently or temporarily
transfer the original mammograms and copies of the patient's reports to a
medical institution, a physician, or to the patient directly.
(C)
If the medical records are permanently forwarded, the receiving
institution or physician shall maintain and become responsible for the original
film until the fifth or tenth anniversary, as specified in subparagraph (A)
of this paragraph.
(5)
Mammographic image identification. Each mammographic image
shall have the following information indicated on it in a permanent, legible
manner and placed so as not to obscure anatomic structures:
(A)
name of patient and an additional patient identifier;
(B)
date of examination;
(C)
view and laterality (this information shall be placed on
the image in a position near the axilla);
(D)
facility name and location (at a minimum the location shall
include city, state, and zip code);
(E)
technologist identification;
(F)
cassette/screen identification; and
(G)
mammography machine identification if there is more than
one machine in the facility.
(6)
Information shall also be maintained for each clinical
image by utilizing a label on each film, recording on the film jacket, or
maintaining a log or other means. The information shall include, but is not
limited to, compressed breast thickness or degree of compression, and kVp.
(u)
Quality assurance - general. Each registrant shall establish
and maintain a written quality assurance program to ensure the safety, reliability,
clarity, and accuracy of mammography services performed at the mammography
facility, including corrective actions to be taken if images are of poor quality.
(1)
Responsible individuals. Responsibility for the quality
assurance program and for each of its elements shall be assigned to individuals
who are qualified for their assignments and who shall be allowed adequate
time to perform these duties.
(A)
Lead interpreting physician. The registrant shall identify
a lead interpreting physician who shall have the general responsibility of:
(i)
ensuring that the quality assurance program meets all requirements
of this subsection and subsections (v) and (w) of this section;
(ii)
reviewing and documenting the technologists' quality control
test results at least every three months or more frequently if consistency
has not yet been achieved;
(iii)
reviewing the physicists' results within 60 days of the
receipt of the results or more frequently when needed; and
(iv)
assigning and determining the individual's qualifications
to perform the quality assurance tasks in subparagraphs (B) - (D) of this
paragraph.
(B)
Interpreting physicians. All interpreting physicians interpreting
mammograms for the registrant shall:
(i)
follow the registrant's procedures for corrective action
when the images they are asked to interpret are of poor quality. These procedures
shall be included in the facility's operating and safety procedures; and
(ii)
participate in the medical outcomes audit program.
(C)
Medical physicist. Each registrant shall use the services
of a licensed medical physicist to survey mammography equipment and oversee
the equipment-related quality assurance practices of the facility. At a minimum,
the medical physicist shall be responsible for performing the surveys and
the mammography equipment evaluations and providing the facility with the
reports described in subsection (v)(10) and (11) of this section.
(D)
Quality control technologist. The quality control technologist,
designated by the lead interpreting physician, shall ensure performance of
the items designated in subsection (v)(1) - (4), (7) - (9), (12), and (14)
of this section. If other personnel are assigned the quality assurance tasks
in accordance with subparagraph (A)(iv) of this paragraph, the quality control
technologist shall insure that the requirements of subsection (v)(1) - (4),
(7) - (9), (12), and (14) of this section are met.
(2)
Quality assurance records. The lead interpreting physician,
quality control technologist, and medical physicist shall ensure that records
concerning mammography technique and procedures, quality control (include
monitoring data, corrective actions, and the effectiveness of the corrective
actions), safety, protection, and employee qualifications to meet assigned
quality assurance tasks are properly maintained and updated. These quality
control records shall be kept for each test specified in subsections (v) and
(w) of this section, in accordance with subsection (ff) of this section.
(v)
Quality assurance - equipment. Registrants with screen-film
systems shall perform the following quality control tests at the intervals
specified. In addition to the intervals specified in paragraphs (4)(B) and
(5)(H) of this subsection, the tests shall be performed prior to initial use.
(1)
Daily quality control tests. Film processors used to develop
mammograms shall be adjusted and maintained to meet the technical development
specifications for the mammography film in use. A processor performance test
shall be completed and the results charted on each day that clinical films
are processed before any clinical films are processed that day.
(A)
Processor performance test. Using mammography film used
clinically at the facility, sensitometer tests shall include assessment of
the following:
(i)
base plus fog density that shall be within plus 0.03 of
the established operating level;
(ii)
mid-density that shall be within plus or minus 0.15 of
the established operating level; and
(iii)
density difference that shall be within plus or minus
0.15 of the established operating level.
(B)
Film processors being used for mammography at multiple
locations, such as a mobile service operation, shall be subject to the requirements
of this paragraph.
(C)
Film processors utilized for mammography shall be adjusted
to and operated at the specifications recommended by the mammographic film
manufacturer, or at other settings such that the sensitometric performance
is at least equivalent.
(D)
Each registrant shall utilize the same film processor for
clinical and phantom images. Clinical images shall be processed within an
interval not to exceed 24 hours from the time the first clinical image is
taken. Facilities utilizing batch processing shall do the following:
(i)
use a container to transport clinical images that will
protect the film from exposure to light and radiation; and
(ii)
maintain a log to include each patient name and unique
identification number, date, and time of the first exam of each batch, and
date and time of batch development.
(2)
Weekly quality control tests. These tests shall be performed
at an interval no greater than seven days. If mammography is not being performed
on the date the test is due and more than seven days have past since the last
test, the tests shall be performed prior to resuming mammography. An image
quality evaluation test, using an FDA-accepted phantom, shall meet the following
parameters.
(A)
The optical density of the film at the center of an image
of a standard FDA-accepted phantom shall be at least 1.20 when exposed under
a typical clinical condition and shall not change by more than plus or minus
0.20 from the established operating level.
(B)
The density difference between the background of the phantom
and an added test object, used to assess image contrast, shall be measured
and shall not vary by more than plus or minus 0.05 from the established operating
level.
(C)
The phantom image shall be made on the standard mammographic
film in use at the facility with techniques used for clinical images of a
standard breast. The phantom image shall meet the requirements in subparagraphs
(A) and (B) of this paragraph and clause (i) of this subparagraph. No mammograms
shall be taken on patients if any of these minimums are not met.
(i)
The mammographic machine shall be capable of producing
images of the mammographic phantom in accordance with the phantom image scoring
protocol in subsection (hh)(4) of this section or paragraph (7) of this subsection.
(ii)
Each phantom image and a record of the evaluation of that
image shall be maintained at the location where the mammography image was
produced or with the radiographic equipment for mobile service operations.
(3)
Quarterly quality control tests. These tests shall be performed
within the calendar quarter at an interval not to exceed 90 days.
(A)
Fixer retention in film. The residual fixer shall be no
more than 5 micrograms per square cm.
(B)
Repeat analysis. A repeat analysis on clinical images repeated
or rejected shall be performed, analyzed, and documented. The total repeat
or reject rate shall not exceed 5.0%. If the total repeat or reject rate changes
from the previously determined rate by more than 2.0% of the total films included
in the analysis, the reason(s) for the change shall be determined. Corrective
action shall be taken and documented if the total repeat or reject rate for
the facility exceeds 5.0% or changes from the previously determined rate by
more than 2.0% of the total films included in the analysis. Test films, cleared
films, or film processed as a result of exposure of a film bin are not to
be included in the count for repeat analysis. Films included in the repeat
analysis are not required to be kept after completion of the analysis.
(4)
Semiannual quality control tests. These tests shall be
performed at an interval not to exceed six months.
(A)
Darkroom fog. The optical density attributable to darkroom
fog shall not exceed 0.05 when a mammography film of the type used in the
facility, which has a mid-density of no less than 1.2 OD, is exposed to typical
darkroom conditions for two minutes while such film is placed on the counter
top, emulsion side up. If the darkroom has a safelight used for mammography
film, it shall be on during this test.
(B)
Screen-film contact. Testing for screen-film contact shall
be conducted using 40 mesh copper screen. The entire area of the cassette
that may be clinically exposed shall be tested. This shall include all cassettes
used for mammography in the facility.
(C)
Compression device performance. The maximum compression
force for the initial power drive shall be between 25 pounds and 45 pounds.
The system shall be capable of compressing the breast with a force of at least
25 pounds and shall be capable of maintaining this compression for at least
15 seconds.
(5)
Annual quality control tests. These tests shall be performed
at an interval not to exceed 14 months.
(A)
Automatic exposure control performance. The AEC shall be
capable of maintaining film optical density within plus or minus 0.15 of the
mean optical density when thickness of a homogeneous material is varied over
a range of 2 to 6 cm and the kVp is varied appropriately for such thicknesses
over the kVp range and in the AEC mode used clinically in the facility.
(B)
Kilovoltage peak accuracy and reproducibility. At the most
commonly used clinical settings of kVp, the coefficient of variation of reproducibility
of the kVp shall be equal to or less than 0.02. The kVp shall be accurate
to within plus or minus 5.0% of the indicated or selected kVp at the following:
(i)
the lowest clinical kVp that can be measured by a kVp test
device;
(ii)
the most commonly used clinical kVp; and
(iii)
the highest available clinical kVp.
(C)
Focal spot condition. Facilities shall evaluate focal spot
condition by determining the system resolution as follows.
(i)
Each system used for mammography, in combination with the
mammography screen-film combination used in the facility, shall provide a
minimum resolution of 11 cycles/millimeter (mm) (line-pairs/mm) when a high
contrast resolution bar test pattern is oriented with the bars perpendicular
to the anode-cathode axis, and a minimum resolution of 13 line-pairs/mm when
the bars are parallel to that axis.
(ii)
The bar pattern shall be placed 4.5 cm above the breast
support surface, centered with respect to the chest wall edge of the image
receptor, and with the edge of the pattern within 1 cm of the chest wall edge
of the image receptor.
(iii)
When more than one target material is provided, the measurement
in clause (i) of this subparagraph shall be made using the appropriate focal
spot for each target material.
(iv)
When more than one SID is provided, the test shall be
performed at the SID most commonly used clinically.
(v)
Test kVp shall be set at the value used clinically by the
facility for a standard breast and shall be performed in the AEC mode, if
available. If necessary, a suitable absorber may be placed in the beam to
increase exposure times. The screen-film cassette combination used by the
facility shall be used to test for this requirement and shall be placed in
the normal location used for clinical procedures.
(D)
Beam quality and half-value layer (HVL). The HVL shall
meet the specifications of Title 21, CFR, §1020.30(m)(l) for the minimum
HVL. These values, extrapolated to the mammographic range, are shown as follows.
This test is performed using the clinical kVp on the standard breast. Values
not shown in Table I may be determined by linear interpolation or extrapolation.
Figure: 25 TAC §289.230(v)(5)(D)
(E) Breast entrance air kerma and AEC reproducibility. The
coefficient of variation for both air kerma and mAs shall not exceed 0.05.
(F) Dosimetry. The average glandular dose delivered during
a single craniocaudal view of an FDA accepted phantom simulating a standard
breast shall not exceed 3.0 milligray (mGy) (0.3 rad) per exposure.
(G) X-ray field/light field/image receptor/compression paddle
alignment. All systems shall meet the following.
(i)
All systems shall have beam-limiting devices that allow
the entire chest wall edge of the x-ray field to extend to the chest wall
edge of the image receptor and provide means to assure that the x-ray field
does not extend beyond any edge of the image receptor by more than 2.0% of
the SID.
(ii)
If a light field that passes through the x-ray beam limitation
device is provided, it shall be aligned with the x-ray field so that the total
of any misalignment of the edges of the light field and the x-ray field along
either the length or the width of the visually defined field at the plane
of the breast support surface shall not exceed 2.0% of the SID.
(iii)
The chest wall edge of the compression paddle shall not
extend beyond the chest wall edge of the image receptor by more than 1.0%
of the SID when tested with the compression paddle placed above the breast
support surface at a distance equivalent to standard breast thickness. The
shadow of the vertical edge of the compression paddle shall not be visible
on the image.
(H)
Uniformity of screen speed. Uniformity of screen speed
of all the cassettes in the facility shall be tested and the difference between
the maximum and minimum optical densities shall not exceed 0.30. Screen artifacts
shall also be evaluated during this test.
(I)
System artifacts. System artifacts shall be evaluated with
a high-grade, defect-free sheet of homogeneous material large enough to cover
the mammography cassette and shall be performed for all cassette sizes used
in the facility using a grid appropriate for the cassette size being tested.
System artifacts shall also be evaluated for all available focal spot sizes
and target filter combinations used clinically.
(J)
Radiation output. The system shall be capable of producing
a minimum output of 7.0 mGy air kerma per second (800 milliroentgen (mR) per
second) when operating at 28 kVp in the standard mammography mode at any SID
where the system is designed to operate. The system shall be capable of maintaining
the required minimum radiation output averaged over a 3.0 second period.
(K)
Decompression. If the system is equipped with a provision
for automatic decompression after completion of an exposure or interruption
of power to the system, the system shall be tested to confirm that it provides
the following:
(i)
an override capability to allow maintenance of compression;
(ii)
a continuous display of the override status; and
(iii)
a manual emergency compression release that can be activated
in the event of power or automatic release failure.
(L)
The technique settings used for subparagraph (F) of this
paragraph and paragraph (2) of this subsection shall be those used by the
facility for its clinical images of a standard breast.
(6)
Densitometer and sensitometer. The calibration of the densitometer
and sensitometer must be in accordance with the manufacturer's specifications.
(7)
Quality control tests - other modalities. For systems with
image receptor modalities other than screen-film, the quality assurance program
shall be substantially the same as the quality assurance program recommended
by the image receptor manufacturer, except that the maximum allowable dose
shall not exceed the maximum allowable dose for screen-film systems in paragraph
(5)(F) of this subsection.
(8)
Mobile service operation. The registrant shall verify that
mammography machines used to produce mammograms at more than one location
meet the requirements in paragraphs (1) - (7) of this subsection. In addition,
at each examination location, before any examinations are conducted, the registrant
shall verify satisfactory performance of the mammography machines by using
a test method that establishes the adequacy of the image quality produced
by the machine. Processor performance shall be in accordance with paragraph
(1) of this subsection.
(9)
Use of test results. After completion of the tests specified
in paragraphs (1) - (8) of this subsection, the following shall occur.
(A)
The registrant shall compare the test results to the corresponding
specified action limits; or, for nonscreen-film modalities, to the manufacturer's
recommended action limits; or for post-move, pre-examination testing of mobile
mammography machines, to the limits established in the test method used by
the facility.
(B)
Components of the mammography system that fail quality
assurance tests shall have corrective actions as indicated in the following.
(i)
If components in subclause (I) and (II) of this clause
fail, corrective action shall be taken before any mammography films are processed:
(I)
paragraph (1) of this subsection describing processor quality
control; and
(II)
paragraph (4)(A) of this subsection describing darkroom
fog;
(ii)
If components in subclause (I) - (VI) of this clause fail,
corrective action shall be taken before any mammography examinations are performed:
(I)
paragraph (2) of this subsection describing phantom image
quality;
(II)
paragraph (4)(B) of this subsection describing screen-film
contact;
(III)
paragraph (4)(C) of this subsection describing compression
device performance;
(IV)
paragraph (5)(F) of this subsection describing dosimetry;
(V)
paragraph (7) of this subsection describing quality control
tests of other modalities; and
(VI)
paragraph (8) of this subsection describing quality control
tests for mobile mammography machines.
(iii)
If components in the remaining quality assurance tests
in subsection (v) of this section fail, corrective action shall be taken within
30 days of the test date.
(C)
Documentation of the tests and the corrective actions described
in subparagraph (B) of this paragraph shall be maintained in accordance with
subsection (ff) of this section.
(10)
Surveys. At least once a year, each facility shall undergo
a survey by a medical physicist or by an individual under the direct supervision
of a medical physicist.
(A)
At a minimum, this survey shall include the following:
(i)
performance of tests to ensure that the facility meets
the quality assurance requirements of the weekly phantom image quality test
described in paragraph (2) of this subsection, the annual tests described
in paragraph (5) of this subsection, and if applicable, quality control tests
as described for other modalities in paragraph (7) of this subsection and
for mobile service operations as described in paragraph (8) of this subsection;
and
(ii)
evaluation of the adequacy of the results of all tests
conducted by the facility as well as written documentation of any corrective
actions taken and their results in accordance with paragraphs (1) - (4) of
this subsection, and, if applicable, paragraphs (7) and (8) of this subsection.
(B)
The medical physicist shall provide a written survey report
to the facility within 30 days of the date of the survey. The report shall
include a summary of the tests performed by the medical physicist in subparagraph
(A)(i) of this paragraph and the review of the tests performed by the facility
in subparagraph (A)(ii) of this paragraph. The report shall also contain recommendations
for any required corrective actions.
(C)
If the following tests indicate deficiencies, the physicist
shall give a preliminary oral or written report to the facility within 72
hours of the survey:
(i)
processor quality control in accordance with paragraph
(9)(B)(i)(I) of this subsection;
(ii)
phantom images, screen-film contact, compression device
performance, or dosimetry in accordance with paragraph (9)(B)(ii)(I) - (IV)
of this subsection;
(iii)
quality control tests for other modalities, if applicable,
in accordance with paragraph (9)(B)(ii)(V) of this subsection; or
(iv)
quality control tests for mobile mammography machines,
if applicable, in accordance with paragraph (9)(B)(ii)(VI) of this subsection.
(D)
The survey report shall be dated and signed by the medical
physicist performing or supervising the survey. If the survey was performed
entirely or in part by another individual under the direct supervision of
the medical physicist, that individual and the part of the survey that individual
performed shall also be identified in the survey.
(E)
The survey report shall be maintained by the registrant
in accordance with subsection (ff) of this section.
(11)
Mammography equipment evaluations. Additional evaluations
of mammography machines or image processors shall be conducted whenever a
new mammography machine or processor is installed, a mammography machine or
processor is disassembled and reassembled at the same or a new location, major
components of mammography machine are changed or repaired, or a processor
is overhauled or reconditioned. These evaluations shall be used to determine
whether the new or changed equipment meets the requirements of applicable
standards in this subsection and subsection (s) of this section.
(A)
All problems shall be corrected before the new or changed
equipment is put into service for examinations or film processing.
(B)
The mammography equipment evaluation and dosimetry shall
be performed by a medical physicist or by an individual under the direct supervision
of a medical physicist.
(12)
Facility cleanliness. The registrant shall establish and
implement adequate protocols for maintaining darkroom, screen, and view box
cleanliness and shall document that all cleaning procedures are performed
at the frequencies specified in the protocols.
(13)
Calibration of air kerma measuring instruments. Instruments
used by medical physicists in their annual survey to measure the air kerma
or air kerma rate from a mammography machine shall be calibrated at least
once every two years and each time the instrument is repaired. The instrument
calibration must be traceable to a national standard and calibrated with an
accuracy of plus or minus 6.0% (95% confidence level) in the mammography energy
range.
(14)
Infection control. Facilities shall establish and comply
with a system specifying procedures to be followed by the facility for cleaning
and disinfecting mammography equipment after contact with blood or other potentially
infectious materials. This system shall specify the methods for documenting
facility compliance with the infection control procedures established and
shall:
(A)
comply with all applicable federal, state, and local regulations
pertaining to infection control; and
(B)
comply with the manufacturer's recommended procedures for
the cleaning and disinfection of the mammography equipment used in the facility;
or
(C)
if adequate manufacturer's recommendations are not available,
comply with generally accepted guidance on infection control, until such recommendations
become available.
(w)
Quality assurance - mammography medical outcomes audit.
Each registrant shall establish and maintain a mammography medical outcomes
audit program to follow-up positive mammographic assessments and to correlate
pathology results with the interpreting physician's findings. This program
shall be designed to ensure the reliability, clarity, and accuracy of the
interpretation of mammograms.
(1)
General requirements. Each registrant shall establish a
system to collect and review outcome data for all mammograms performed, including
follow-up on the disposition of all positive mammograms and correlation of
pathology results with the interpreting physician's mammography report. Analysis
of these outcome data shall be made individually and collectively for all
interpreting physicians at the facility. In addition, any cases of breast
cancer among women imaged at the facility that subsequently become known to
the facility shall prompt the facility to initiate follow-up on surgical and/or
pathology results and review of the mammograms taken prior to the diagnosis
of a malignancy.
(2)
Frequency of audit analysis. The facility's first audit
analysis shall be initiated no later than 12 months after the date the facility
becomes certified or 12 months after April 28, 1999, whichever date is the
latest. This audit analysis shall be complete within an additional 12 months
to permit completion of diagnostic procedures and data collection. Subsequent
audit analyses will be conducted at least once every 12 months. These shall
be maintained in accordance with subsection (ff) of this section.
(3)
Reviewing interpreting physician. Each lead interpreting
physician or an interpreting physician designated by the lead interpreting
physician shall review the medical outcomes audit data at least once every
12 months. This individual shall analyze the results of the audit and shall
be responsible for the following:
(A)
recording the dates of the audit period(s);
(B)
documenting the results;
(C)
notifying other interpreting physicians of their results
and the registrant's aggregate results; and
(D)
documenting any follow up actions and the nature of the
follow up.
(x)
Mammographic procedure and techniques for mammography of
patients with breast implants. Each registrant shall have a procedure to inquire
whether or not the patient has breast implants prior to the mammographic exam.
Except where contraindicated, or unless modified by a physician's directions,
patients with breast implants shall have mammographic views to maximize the
visualization of breast tissue.
(y)
Complaints. Each accredited facility shall do the following:
(1)
establish a written procedure for collecting and resolving
consumer complaints;
(2)
maintain a record of each serious complaint received by
the facility in accordance with subsection (ff) of this section; and
(3)
report unresolved serious complaints to the facility's
FDA-approved accreditation body within 30 days of receiving the complaint.
(z)
Clinical image quality. Clinical images produced by any
certified facility must continue to comply with the standards for clinical
image quality established by that facility's accreditation body.
(aa)
Additional mammography review, targeted clinical reviews,
and patient notification.
(1)
If the agency certifying body believes that mammography
quality at a facility may have been compromised and presents a serious risk
to human health, the facility shall provide clinical images and other relevant
information, as specified by the agency certifying body, for review by the
FDA-approved accreditation body.
(2)
If the agency certifying body determines that mammography
quality at a facility has been compromised and presents a serious risk to
human health, the facility shall provide clinical images and other relevant
information, as specified by the agency certifying body, for review by the
FDA-approved accreditation body. The agency certifying body may require such
facility to notify patients who received mammograms, and their referring physicians.
The notification shall include the deficiencies presenting such risk, the
potential consequences to the patient, appropriate remedial measures, and
such other relevant information as the agency certifying body may require.
Such notification shall occur within a time frame and in a manner specified
by the agency.
(3)
The agency certifying body, the agency accreditation body
or another FDA-approved accreditation body, or the FDA may request a targeted
clinical image review due to, but not limited to, serious complaints or severe
items of non-compliance.
(bb)
Requirements for machines used exclusively for interventional
breast radiography. Machines used exclusively for interventional breast radiography,
including mobile service operations, are not included in the definition of
mammography systems. These machines are not required to be accredited or to
receive certification by the agency certifying body in accordance with 21
CFR, Part 900.11. However, each facility using such machines shall apply for
and receive a certification from the agency. The facility shall comply with
the following:
(1)
purpose and scope in accordance with subsections (a) and
(b) of this section;
(2)
applicable definitions in subsection (c) of this section;
(3)
prohibitions in accordance with subsection (d)(2) and (3)
of this section;
(4)
exemptions in accordance with subsection (e)(2), (3), and
(5) of this section;
(5)
certification requirements in accordance with subsection
(f)(2) - (4) and (5)(B) of this section and the requirement to submit a medical
physicist's survey in accordance with paragraph (13) of this subsection;
(6)
issuance of certification and specific terms and conditions
of certification in accordance with subsections (g)(1) and (1)(B), (2), (3),
and (m) of this section;
(7)
responsibilities of a registrant in accordance with subsection
(n)(1), (2)(A), (D) and (E), and (4) - (6) of this section;
(8)
expiration, termination, modification and revocation of
certification in accordance with subsections (l), (p), and (q) of this section;
(9)
renewal of certification as follows:
(A)
the registrant shall file an application for renewal of
certification in accordance with subsection (f)(2) - (4) and (5)(B) of this
section and submit a medical physicist's survey in accordance with paragraph
(13) of this subsection; and
(B)
if a registrant files an application in proper form at
least 30 days before the existing certification expires, such existing certification
shall not expire until the application status has been determined by the agency
certifying body;
(10)
personnel requirements for a general certificate, medical
radiologic technologist in accordance with the Medical Radiologic Technologist
Certification Act, Texas Occupational Code, Chapter 601;
(11)
personnel requirements for medical physicists in accordance
with subsection (r)(3) of this section;
(12)
requirement to have a written quality assurance program
to ensure the safety, reliability, clarity, and accuracy of services performed
at the facility, including corrective actions to be taken if images are of
poor quality;
(13)
requirement to have a medical physicist perform an annual
survey of AEC, kVp, focal spot condition, HVL, and dosimetry tests in accordance
with subsection (v)(5)(A) - (F) of this section. The medical physicist shall
provide a preliminary oral or written report of deficiencies within 72 hours
of the survey if it involves dosimetry. The medical physicist shall prepare
a written report for the facility within 30 days of the date of the survey
to include the following:
(A)
a summary of the tests in the annual survey with recommendations
for corrective actions; and
(B)
date and signature of the medical physicist performing
or supervising the survey. If the survey was performed entirely or in part
by another individual under the direct supervision of the medical physicist,
that individual and the part of the survey that individual performed shall
also be identified in the survey;
(14)
the requirement to correct deficiencies indicated in the
test results for dosimetry in accordance with subsection (v)(9)(B)(ii)(IV)
of this section before any further examinations are performed;
(15)
operating and safety procedures in accordance with subsection
(ee)(1) of this section;
(16)
occupational dose limits and personnel monitoring in accordance
with §289.231 of this title;
(17)
provision of a technique chart in accordance with subsection
(ee)(2) of this section;
(18)
the requirement to maintain receipt, transfer, disposal,
calibration, and maintenance records in accordance with subsection (ee)(3)
and (8) of this section;
(19)
requirement to have a viewing system in accordance with
subsection (ee)(4) of this section;
(20)
requirement to prevent exposure of individuals other than
the patient in accordance with subsection (ee)(5) of this section;
(21)
maintenance of applicable records in subsection (ff) of
this section;
(22)
inspection requirements in accordance with subsection
(gg) of this section, except for subsection (gg)(1) of this section; and
(23)
equipment requirements in accordance with §289.227(h)
of this title (relating to Use of Radiation Machines in the Healing Arts).
(cc)
Self-referral mammography. Any person proposing to conduct
a self-referral mammography program shall not initiate such a program without
prior approval of the agency. When requesting such approval, that person shall
submit the following information:
(1)
the number and type of views (or projections);
(2)
the age of the population to be examined and the frequency
of the exam following established, nationally recognized criteria, such as
those of the American Cancer Society, American College of Radiology (ACR),
or the National Council on Radiation Protection and Measurements;
(3)
written procedures to include methods of:
(A)
advising patients and private physicians of the results
of the mammography examination in accordance with subsection (t)(2) of this
section;
(B)
follow-up with patients and physicians in accordance with
subsection (t)(3) of this section; and
(C)
recommending to patients who do not have a physician means
of selecting a physician; and
(4)
methods for educating mammography patients in breast self-examination
techniques and on the necessity for follow-up by a physician.
(dd)
Medical research and investigational devices.
(1)
Any research using radiation producing devices on humans
must be approved by an IRB as required by Title 45, CFR, Part 46 and Title
21, CFR, Part 56. The IRB must include at least one licensed physician to
direct any use of radiation in accordance with §289.231(b) of this title.
(2)
Facilities with mammography machines with investigational
device exemptions that are involved in clinical studies must comply with primary
regulations that govern the conduct of clinical studies and that apply to
the manufacturers, sponsors, clinical investigators, institutional review
boards, and the medical device. These regulations include the following:
(A)
21 Code of Federal Regulations (CFR), Part 812,
Investigational Device Exemptions
;
(B)
21 CFR, Part 50,
Protection of
Human Subjects
;
(C)
21 CFR, Part 56,
Institutional
Review Boards
;
(D)
21 CFR, Part 54,
Financial Disclosure
by Clinical Investigators
; and
(E)
21 CFR, Part 821, Subpart C,
Design Controls of the Quality System Regulation
.
(ee)
Other operating procedures.
(1)
Operating and safety procedures. Each registrant shall
have and implement written operating and safety procedures that shall be made
available to each individual operating x-ray equipment, including any restrictions
of the operating technique required for the safe operation of the particular
system. These procedures shall include, but are not limited to, the items
in subsection (hh)(3) of this section.
(2)
Technique chart. A chart or manual shall be provided or
electronically displayed in the vicinity of the control panel of each machine
that specifies technique factors to be utilized versus patient's anatomical
size. The technique chart shall be used by all operators.
(3)
Receipt, transfer, and disposal of mammography machines.
Each registrant shall maintain records showing the receipt, transfer, and
disposal of mammographic machines. These records shall include the date of
receipt, transfer, or disposal; the name and signature of the individual making
the record; and the manufacturer's model and serial number from the control
panel of the mammographic machine. Records shall be maintained in accordance
with subsection (ff) of this section for inspection by the agency.
(4)
Viewing system. Windows, mirrors, closed circuit television,
or an equivalent system shall be provided to permit the operator to continuously
observe the patient during irradiation. The operator shall be able to maintain
verbal, visual, and aural contact with the patient.
(5)
Exposure of individuals other than the patient. Only the
staff and ancillary personnel required for the medical procedure or training
shall be in the room during the radiation exposure unless such individual's
assistance is required.
(6)
Protective devices. Protective devices shall be utilized
when required, as in paragraph (7) of this subsection.
(A)
Protective devices shall be of no less than 0.25 mm lead
equivalent material.
(B)
Protective devices, including aprons, gloves, and shields
shall be checked annually for defects such as holes, cracks, and tears. These
checks may be performed by the registrant by visual or tactile means, or x-ray
imaging. If a defect is found, protective devices shall be replaced or removed
from service until repaired. A record of this test shall be made and maintained
by the registrant in accordance with subsection (ff) of this section for inspection
by the agency.
(7)
Holding of patient or image receptor.
(A)
When a patient or image receptor must be held in position
during radiography, mechanical supporting or restraining devices shall be
used when the exam permits.
(B)
If a patient or image receptor must be held by an individual
during an exposure, that individual shall be protected with appropriate shielding
devices described in paragraph (6) of this subsection.
(C)
The registrant's written operating and safety procedures
required by paragraph (1) of this subsection shall include the following:
(i)
a list of circumstances in which mechanical holding devices
cannot be routinely utilized; and
(ii)
a procedure used for selecting an individual to hold or
support the patient or image receptor.
(D)
In those cases where the patient must hold the image receptor,
any portion of the body other than the area of clinical interest struck by
the useful beam shall be protected by not less than 0.25 mm lead equivalent
material.
(8)
Calibration, maintenance, and modifications. Each registrant
shall maintain records showing calibrations, maintenance, and modifications
performed on each mammographic machine. These records shall include the date
of the calibration, maintenance, or modification performed; the name of the
individual making the record; and the manufacturer's model and serial number
of the control panel of the mammographic machine. These records shall be maintained
in accordance with subsection (ff) of this section.
(ff)
Record requirements. Records required by this section
shall be maintained for inspection by the agency in accordance with paragraph
(3) of this subsection. Records may be maintained electronically in accordance
with §289.231(ff)(3) of this title.
(1)
Records for mammography machines authorized for mobile
service operations.
(A)
Copies of the following shall be kept with mammography
machines authorized for mobile services:
(i)
operating and safety procedures in accordance with subsection
(ee)(1) of this section;
(ii)
medical radiologic technologists' credentials;
(iii)
current quality control records for at least the last
90 calendar days for on-board processors in accordance with subsection (v)(1)
of this section;
(iv)
current §289.203 of this title, §289.226 of
this title, §289.230 of this title, §289.231 of this title, and §289.234
of this title if accredited by the agency accreditation body;
(v)
copy of certification;
(vi)
certification of inspection in accordance with subsection
(gg)(5) of this section;
(vii)
notice of failure from last inspection in accordance
with subsection (gg)(6) of this section, if applicable; and
(viii)
copy of mammography accreditation.
(B)
Copies of all other records required by this section shall
be maintained at a specified location.
(2)
Records required at separate authorized use locations.
Copies of the following shall be kept at each separate authorized use location:
(A)
credentials for interpreting physicians operating at that
location in accordance with subsection (r)(1) of this section;
(B)
credentials for medical radiologic technologists operating
at that location in accordance with subsection (r)(2) of this section;
(C)
credentials for medical physicists operating at that location
in accordance with subsection (r)(3) of this section;
(D)
continuing education and experience records for interpreting
physicians, medical radiologic technologists, and medical physicists operating
at that location in accordance with subsection (r)(1)(C), (2)(C), and (3)(C)
of this section;
(E)
mandatory training records for interpreting physicians,
medical radiologic technologists, and medical physicists operating at that
location in accordance with subsection (r)(1)(E), (2)(E), and (3)(E) of this
section, if applicable;
(F)
current physicist annual survey of the mammography system;
(G)
current §289.203 of this title, §289.226 of this
title, §289.230 of this title, §289.231 of this title, and §289.234
of this title if accredited by the agency accreditation body;
(H)
copy of certification;
(I)
quality assurance program in accordance with subsections
(u), (v), and (w) of this section;
(J)
quality control records in accordance with subsection (u)(2)
of this section;
(K)
operating and safety procedures in accordance with subsection
(ee)(1) of this section;
(L)
records of receipts, transfers, and disposal in accordance
with subsection (ee)(3) of this section;
(M)
calibration, maintenance, and modification records in accordance
with subsection (ee)(8) of this section;
(N)
certification of inspection in accordance with subsection
(gg)(5) of this section;
(O)
notification of failure in accordance with subsection (gg)(6)
of this section, if applicable;
(P)
records of notification of patients in accordance with
subsection (gg)(10) this section; and
(Q)
copy of mammography accreditation.
(3)
Time requirements for record keeping. Time requirements
for record keeping shall be according to the following chart.
Figure: 25 TAC §289.230(ff)(3)
(gg)
Inspections. In addition to the requirements of §289.231(kk)
of this title, the following applies to inspections of mammography systems.
(1)
The agency may inspect each mammography system that receives
a certification in accordance with this chapter not later than the 60th day
after the date the certification is issued.
(2)
The agency may inspect, at least once annually, each mammography
system that receives a certification.
(3)
To protect the public health, the agency may conduct more
frequent inspections than required by this subsection.
(4)
The agency may make reasonable attempts to coordinate inspections
in this section with other inspections required in accordance with this chapter
for the facility where the mammography system is used.
(5)
After each satisfactory inspection, the agency shall issue
a certificate of inspection for each mammography system inspected. The certificate
of inspection shall be posted at a conspicuous place on or near the place
where the mammography system is used. The certificate of inspection may include
the following:
(A)
specific identification of the mammography system inspected;
(B)
the name and address of the facility where the mammography
system was used at the time of the inspection; and
(C)
the date of the inspection.
(6)
Any severity level I violation involving a mammography
system, found by the agency, in accordance with §289.205 of this title,
constitutes grounds for posting notice of failure of the mammography system
to satisfy agency requirements.
(A)
Notification of such failure shall be posted:
(i)
on the mammography machine at a conspicuous place if the
violation is machine-related; or
(ii)
near the place where the mammography system practices
if the violation is personnel-related; and
(iii)
in a sufficient number of places to permit the patient
to observe the notice.
(B)
The notice of failure shall remain posted until the facility
is authorized to remove it by the agency. A facility may post documentation
of corrections of the violations submitted to the agency along with the notice
of failure until approval to remove the notice of failure is received from
the agency.
(7)
Facilities that receive a severity level I violation shall
notify patients on whom the facility performed a mammogram during the 30 days
preceding the date of the inspection that revealed the failure. The facility
shall:
(A)
inform the patient that the mammography system failed to
satisfy the agency certifying body's standards;
(B)
recommend that the patient have another mammogram performed
at a facility with a certified mammography system; and
(C)
list the three facilities closest to the original testing
facility that have a certified mammography system.
(8)
In addition to the requirements of paragraph (7) of this
subsection, the agency may require a facility to notify a patient of any other
failure of the facility's mammography system to meet the agency's certification
standards.
(9)
The patient notification shall include the following:
(A)
an explanation of the mammography system failure to the
patient; and
(B)
the potential consequences to the mammography patient.
(10)
The registrant shall make a record of the mammography
patients notified in accordance with paragraphs (7) and (8) of this subsection
for inspection by the agency. The records shall include the name and address
of each mammography patient notified, date of notification, and a copy of
the text sent to the individual. The records shall be maintained in accordance
with subsection (ff) of this section.
(hh)
Appendices.
(1)
Subjects to be included in mammography training for medical
radiologic technologists shall include, but not be limited to, the following:
(A)
breast anatomy and physiology;
(B)
positioning and compression;
(C)
quality assurance/quality control techniques;
(D)
imaging of patients with breast implants; and
(E)
at least eight hours of training in each mammography modality
to be used by the technologist in performing mammography exams.
(2)
Subjects to be included in mammography training for interpreting
physicians shall include, but not be limited to, the following:
(A)
radiation physics, including radiation physics specific
to mammography;
(B)
radiation effects;
(C)
radiation protection; and
(D)
interpretation of mammograms. This shall be under the direct
supervision of a physician who meets the requirements of subsection (r)(1)
of this section.
(3)
Operating and safety procedures. The registrant's operating
and safety procedures shall include, but are not limited to, the following
procedures as applicable:
(A)
posting notices to workers in accordance with §289.203(b)
of this title;
(B)
instructions to workers in accordance with §289.203(c)
of this title;
(C)
notifications and reports to individuals in accordance
with §289.203(d) of this title;
(D)
ordering x-ray exams in accordance with §289.231(b)
of this title;
(E)
occupational dose requirements in accordance with §289.231(m)
of this title;
(F)
personnel monitoring requirements in accordance with §289.231(n)
and (q) of this title;
(G)
posting of a radiation area in accordance with §289.231(x)
and (y) of this title;
(H)
credentialing requirements for lead interpreting physicians,
interpreting physicians, medical radiologic technologists, and medical physicists
in accordance with subsection (r) of this section;
(I)
retention of clinical images in accordance with subsection
(t)(4) of this section;
(J)
quality assurance program in accordance with subsections
(u) - (w) of this section;
(K)
image quality and corrective action for images of poor
quality in accordance with subsection (u)(1)(B)(i) of this section;
(L)
repeat analysis in accordance with subsection (v)(3)(B)
of this section;
(M)
procedures and techniques for mammography patients with
breast implants in accordance with subsection (x) of this section;
(N)
procedure to handle complaints in accordance with subsection
(y) of this section;
(O)
self-referral mammography in accordance with subsection
(cc) of this section;
(P)
use of a technique chart in accordance with subsection
(ee)(2) of this section;
(Q)
exposure of individuals other than the patient in accordance
with subsection (ee)(5) of this section;
(R)
use of protective devices in accordance with subsection
(ee)(6) of this section; and
(S)
holding of patients or image receptors in accordance with
subsection (ee)(7) of this section.
(4)
Phantom image scoring protocol for film-screen modality.
Each of the following object groups are to be scored separately. In order
to receive a passing score on the phantom image, all three test object groups
must pass. A failure in any one of the areas results in a phantom failure.
(A)
Fibers. A score of 4.0 for fibers is required to meet the
evaluation criteria. The diameter size of fibers are 1.56 mm, 1.12 mm, 0.89
mm, 0.75 mm, 0.54 mm, and 0.40 mm. Score the fibers as follows.
(i)
Begin with the largest fiber and move down in size, adding
one point for each full fiber until a score of zero or one half is given.
Stop counting at the first point where you lose visibility of objects.
(ii)
If the entire length of the fiber can be seen and its
location and orientation are correct, that fiber receives a score of one.
(iii)
If at least half, but not all, of the fiber can be seen
and its location and orientation are correct, that fiber receives a score
of one half.
(iv)
If less than one half of a fiber can be seen or if the
location or orientation are incorrect, that fiber receives a score of zero.
(v)
After determining the last fiber to be counted, look at
the overall background for artifacts. If there are background objects that
are fiber-like in appearance and are of equal or greater brightness than the
last visible half or full fiber counted, subtract the last half or full fiber
scored.
(B)
Speck groups. A score of 3.0 for speck groups is required
to meet the evaluation criteria. Diameter sizes of speck groups are 0.54 mm,
0.40 mm, 0.32 mm, 0.24 mm, and 0.16 mm. There are six specks per group. Score
the speck groups as follows.
(i)
Begin with the largest speck group and move down in size
adding one point for each full speck group until a score of one half or zero
is given, then stop.
(ii)
If at least four of the specks in any group are visualized,
the speck group is scored as one.
(iii)
If two or three specks in a group are visualized, the
score for the group is one half.
(iv)
If one speck or no specks from a group are visualized,
the score is zero.
(v)
After determining the last speck group to receive a full
or one-half point, look at the overall background for artifacts. If there
are speck-like artifacts within the insert region of the phantom that are
of equal or greater brightness than individual specks counted in the last
visible half or full speck group counted, subtract the artifact speck from
the observed specks in the last group scored, one by one. Note that the highest
number of speck-like artifacts that can potentially be subtracted is the number
of visible specks that were scored in the last group. Repeat the scoring of
the last visible speck group after these deductions.
(C)
Masses. A score of 3.0 is required to meet the evaluation
criteria. Diameter sizes of masses are 2.00 mm, 1.00 mm, 0.75 mm, 0.50 mm,
and 0.25 mm. Score the masses as follows.
(i)
Begin with the largest mass and add one point for each
full mass observed until a score of one half or zero is assigned.
(ii)
Score one for each mass that appears as a minus density
object in the correct location that can be seen clearly enough to observe
round, circumscribed borders.
(iii)
Score one half if the mass is clearly present in the
correct location, but the borders are not visualized as circular.
(iv)
After determining the last full or half mass to be counted,
look at the overall background for artifacts. If there are background objects
that are mass-like in appearance and are of equal or greater visibility than
the last visible mass, subtract the last full or half point assigned from
the original score.
This agency hereby certifies that the proposal
has been reviewed by legal counsel and found to be within the agency's legal
authority to adopt.
Filed
with the Office of the Secretary of State on June 9, 2006.
TRD-200603127
Cathy Campbell
General Counsel
Department of State Health Services
Earliest possible date of adoption: July 23, 2006
For further information, please call: (512) 458-7111 x6972
25 TAC §289.234
The Executive Commissioner of the Health and Human Services
Commission (commission), on behalf of the Department of State Health Services
(department), proposes new §289.234, concerning mammography accreditation.
BACKGROUND AND PURPOSE
Section 289.230 concerning certification of mammography systems and accreditation
of mammography facilities is being repealed and divided into two new rules,
one concerning certification and the other concerning accreditation. The repeal
and new rules are the result of the agency applying to the United States Food
and Drug Administration (FDA) to become a certifying body for mammography
facilities. The FDA recommended that the current rules be separated into certification
and accreditation rules for clarification, as all mammography facilities in
the state must have certification with the agency, while accreditation may
be with the agency or with the American College of Radiology. The accreditation
requirements will be incorporated into new §289.234, "Mammography Accreditation."
Certification requirements will be located in §289.230 of this title,
"Certification of Mammography Systems and Mammography Machines Used for Interventional
Breast Radiography" that is addressed in a separate preamble.
SECTION-BY-SECTION SUMMARY
The new section adds definitions for agency accreditation body and agency
certifying body to define terms used in this section. Definitions for accreditation
and targeted clinical image review are revised to be consistent with the FDA's
Mammography Quality Standards Act (MQSA) rules. The section contains requirements
for facilities that choose to be accredited with the agency including denials,
suspensions and revocations, and appeals.
FISCAL NOTE
Susan E. Tennyson, Section Director, Environmental and Consumer Safety
Section, has determined that for each year of the first five-year period that
the section will be in effect, there will be no fiscal implications to state
or local governments as a result of enforcing and administering the section
as proposed.
SMALL AND MICRO-BUSINESS IMPACT ANALYSIS
Ms. Tennyson has also determined that there will be no effect on small
businesses or micro-businesses required to comply with the section as proposed.
This was determined by interpretation of the rule that small businesses and
micro-businesses will not be required to alter their business practices in
order to comply with the section. There is no anticipated negative impact
on local employment.
PUBLIC BENEFIT
In addition, Ms. Tennyson has also determined that for each year of the
first five years the section is in effect, the public will benefit from adoption
of the sections. The public benefit anticipated as a result of enforcing or
administering the section is to protect public health and safety by affording
facilities the opportunity to become certified and accredited for mammography
through one agency. Facilities will also have the option of becoming accredited
through the American College of Radiology.
REGULATORY ANALYSIS
The department has determined that this proposal is not a "major environmental
rule" as defined by Government Code, §2001.0225. "Major environmental
rule" is defined to mean a rule the specific intent of which is to protect
the environment or reduce risk to human health from environmental exposure
and that may adversely affect, in a material way, the economy, a sector of
the economy, productivity, competition, jobs, the environment or the public
health and safety of a state or a sector of the state. This proposal is not
specifically intended to protect the environment or reduce risks to human
health from environmental exposure.
TAKINGS IMPACT ASSESSMENT
The department has determined that the proposed new rule does not restrict
or limit an owner's right to his or her property that would otherwise exist
in the absence of government action and, therefore, does not constitute a
taking under Government Code, §2007.043.
PUBLIC COMMENT
Comments on the proposal may be submitted to Cindy Cardwell, Radiation
Group, Policy/Standards/Quality Assurance Unit, Environmental and Consumer
Safety Section, Division of Regulatory Services, Department of State Health
Services, 1100 West 49th Street, Austin, Texas 78756, (512) 834-6770, extension
2239, or by email to cindy.cardwell@dshs.state.tx.us. Comments will be accepted
for 30 days following publication of the proposal in the
Texas Register
.
PUBLIC HEARING
A public hearing to receive comments on the proposal will be scheduled
after publication in the
Texas Register
and
will be held at the Department of State Health Services, Exchange Building,
8407 Wall Street, Austin, Texas 78754. The meeting date will be posted on
the Radiation Control web site (www.dshs.state.tx.us/radiation). Please contact
Cindy Cardwell at (512) 834-6770, extension 2239, or cindy.cardwell@dshs.state.tx.us,
if you have questions.
LEGAL CERTIFICATION
The Department of State Health Services General Counsel, Cathy Campbell,
certifies that the proposed rule has been reviewed by legal counsel and found
to be within the state agencies' authority to adopt.
STATUTORY AUTHORITY
The new section is authorized by Health and Safety Code, §401.051,
which provides the Executive Commissioner of the Health and Human Services
Commission with authority to adopt rules and guidelines relating to the control
of radiation, and Government Code, §531.0055, and Health and Safety Code, §1001.075,
which authorize the Executive Commissioner of the Health and Human Services
Commission to adopt rules and policies necessary for the operation and provision
of health and human services by the department and for the administration
of Health and Safety Code, Chapter 1001.
The new section affects the Health and Safety Code, Chapters 401 and 1001;
and Government Code, Chapter 531.
§289.234.Mammography Accreditation.
(a)
Purpose. This section provides for the accreditation of
mammography facilities. The use of all mammography machines accredited in
accordance with this section shall be by or under the supervision of a physician
licensed by the Texas Medical Board.
(b)
Scope. In addition to the requirements of this section,
all mammography facilities are subject to the requirements of §289.204
of this title (relating to Fees for Certificates of Registration, Radioactive
Material Licenses, Emergency Planning and Implementation, and Other Regulatory
Services), §289.205 of this title (relating to Hearing and Enforcement
Procedures), §289.230 of this title (relating to Certification of Mammography
Systems and Mammography Machines Used for Interventional Breast Radiography),
and §289.231 of this title (relating to General Provisions and Standards
for Protection Against Machine-Produced Radiation). This section does not
apply to an entity under the jurisdiction of the federal government.
(c)
Definitions. The following words and terms, when used in
this section, shall have the following meanings unless the context clearly
indicates otherwise.
(1)
Accreditation--An approval of a mammography machine within
a mammography facility by an accreditation body. A facility may be accredited
by the agency accreditation body or another United States Food and Drug Administration
(FDA)-approved accreditation body.
(2)
Act--Texas Radiation Control Act, Health and Safety Code,
Chapter 401.
(3)
Additional mammography review--At the request of the FDA
or an accreditation body, a review by the accreditation body of clinical images
and other relevant facility information necessary to assess conformation with
the accreditation standards. The reviews include the following:
(A)
additional mammography review with interpretation; or
(B)
additional mammography review without interpretation.
(4)
Adverse event--An undesirable experience associated with
mammography activities within the scope of this section. Adverse events include
but are not limited to:
(A)
poor image quality;
(B)
failure to send mammography reports within 30 days to the
referring physician or in a timely manner to the self-referred patient; and
(C)
use of personnel who do not meet the applicable requirements
in §289.230(r) of this title.
(5)
Agency accreditation body--For the purpose of this section,
the agency as approved by the FDA under Title 21, Code of Federal Regulations
(CFR), Part 900.3(d), to accredit mammography facilities in the State of Texas.
(6)
Agency certifying body--For the purpose of this section,
the agency, as approved by FDA, under Title 21, CFR, Part 900.21, that certifies
facilities within the State of Texas to perform mammography services.
(7)
Certification--An authorization for the use of a mammography
system or the certification of mammography machines used for interventional
breast radiography.
(8)
Clinical image--See the definition for mammogram.
(9)
Consumer--An individual who chooses to comment or complain
in reference to a mammography examination. The individual may be the patient
or a representative of the patient, such as a family member or referring physician.
(10)
Facility--A hospital, outpatient department, clinic, radiology
practice, mobile unit, an office of a physician, or other person that conducts
breast cancer screening or diagnosis through mammography activities, including
the following:
(A)
the operation of equipment to produce a mammogram;
(B)
processing of film;
(C)
initial interpretation of the mammogram; or
(D)
maintaining the viewing conditions for that interpretation.
(11)
FDA-approved accreditation body--An entity approved by
the FDA under Title 21, CFR, Part 900.3(d), to accredit mammography facilities.
(12)
Healing arts--Any system, treatment, operation, diagnosis,
prescription, or practice for the ascertainment, cure, relief, palliation,
adjustment, or correction of any human disease, ailment, deformity, injury,
or unhealthy or abnormal physical or mental condition.
(13)
Image review board--A group of qualified review physicians
and other individuals who review the clinical and phantom images and whose
qualifications have been established by the accreditation body and the accreditation
body's qualifications have been approved by the FDA.
(14)
Interpreting physician--A licensed physician who interprets
mammographic images and who meets the requirements of §289.230(r)(1)
of this title.
(15)
Mammogram--A radiographic image produced through mammography.
(16)
Mammography--The use of x-radiation to produce an image
of the breast that may be used to detect the presence of pathological conditions
of the breast. For the purposes of this section, mammography does not include
radiography of the breast performed as follows:
(A)
during invasive interventions for localization or biopsy
procedures except as specified in §289.230(z) of this title; or
(B)
with an investigational mammography device as part of a
scientific study conducted in accordance with FDA's investigational device
exemption regulations.
(17)
Mammography machine(s)--A unit consisting of components
assembled for the production of x-rays for use during mammography. These include,
at a minimum, the following:
(A)
an x-ray generator;
(B)
an x-ray control;
(C)
a tube housing assembly;
(D)
a beam limiting device; and
(E)
supporting structures.
(18)
Mammography system--A system that includes the following:
(A)
an x-ray machine used as a source of radiation in producing
images of breast tissue;
(B)
an imaging system used for the formation of a latent image
of breast tissue;
(C)
an imaging-processing device for changing a latent image
of breast tissue to a visual image that can be used for diagnostic purposes;
(D)
a viewing device used for the visual evaluation of an image
of breast tissue if the image is produced in interpreting visual data captured
on an image receptor;
(E)
a medical radiologic technologist who performs mammography;
and
(F)
a physician who engages in, and who meets the requirements
of this section relating to the reading, evaluation, and interpretation of
mammograms.
(19)
Medical physicist--An individual who performs surveys
and evaluations of mammographic equipment and facility quality assurance programs
in accordance with this section and who meets the qualifications in §289.230(r)(3)
of this title.
(20)
Medical radiologic technologist (operator of equipment)--An
individual specifically trained in the use of radiographic equipment and the
positioning of patients for radiographic examinations, who performs mammography
examinations in accordance with this section and who meets the qualifications
in §289.230(r)(2) of this title.
(21)
Patient--Any individual who undergoes a mammography examination
in a facility, regardless of whether the person is referred by a physician
or is self-referred.
(22)
Phantom--A test object used to simulate radiographic characteristics
of compressed breast tissue and containing components that radiographically
model aspects of breast disease and cancer.
(23)
Phantom image--A radiographic image of a phantom.
(24)
Radiation machine--For the purposes of this part, radiation
machine also means mammography machine.
(25)
Reinstatement fee--The fee in accordance with §289.204(h)
of this title charged to reinstate an application for a mammography machine
that has been denied accreditation or whose application has been abandoned
in accordance with subsection (h)(3) of this section.
(26)
Review physician--An individual who is qualified to review
clinical images on behalf of the accreditation body. To be qualified, this
individual shall comply with the following:
(A)
meet the interpreting physician requirements of §289.230(r)(1)
of this title;
(B)
be trained and evaluated in the clinical image review process
for the types of clinical images to be evaluated by a review physician by
the accreditation body before designation as a review physician and periodically
thereafter; and
(C)
clearly document findings and reasons for assigning a particular
score to any clinical image and provide information to the facility for use
in improving the attributes for which significant deficiencies were identified.
(27)
Serious adverse event--An adverse event that may significantly
compromise clinical outcomes, or an adverse event for which a facility fails
to take appropriate corrective action in a timely manner.
(28)
Serious complaint--A report of a serious adverse event.
(29)
Survey--An on-site physics consultation and evaluation
of a facility quality assurance program performed by a medical physicist.
(30)
Targeted clinical image review--A review of a minimum
of two sets of "negative" clinical images from a specific date, or date range,
at the request of the agency.
(d)
Accreditation of mammography facilities.
(1)
All mammography facilities shall be accredited by an FDA-approved
accreditation body and shall meet the quality standards in §289.230(r)
- (aa) of this title. In order to qualify for certification in accordance
with §289.230 of this title, new facilities applying to the agency accreditation
body shall receive acceptance of the accreditation application.
(2)
The facility shall submit the following information in
addition to the information required in subsection §289.230(f) of this
title:
(A)
an application for accreditation on forms and in accordance
with accompanying instructions prescribed by the agency accreditation body;
(B)
the appropriate accreditation fee prescribed in §289.204
of this title; and
(C)
evidence that the medical physicist's survey and mammography
equipment evaluation in accordance with §289.230(v)(10) and (11) of this
title was performed within the following time frames:
(i)
no more than six months before the date of the accreditation
application for new facilities seeking accreditation;
(ii)
no more than 14 months before the date of the application
for accreditation for facilities changing accreditation to one issued by the
agency accreditation body; or
(iii)
no more than 14 months before the date of the application
for renewal of accreditation for facilities accredited by the agency accreditation
body.
(3)
Upon notification by the agency accreditation body, each
applicant shall submit clinical and phantom images directly to the image review
board.
(e)
Issuance of accreditation of a mammography facility. An
accreditation document will be issued when the mammography facility meets
the requirements of subsection (d) of this section and §289.204 of this
title and becomes accredited by the agency accreditation body. In order for
an accreditation to be issued, the agency accreditation body must receive
acceptable dose evaluation information from the dosimetry processor and be
notified by the image review board that the applicant met the criteria for
clinical images and phantom images, and dose evaluation.
(f)
Denial or abandonment of an application for accreditation
of mammography facilities.
(1)
Any application for accreditation may be denied by the
agency accreditation body when the applicant fails to meet established criteria
for accreditation in accordance with subsection (d) of this section.
(2)
Before the agency accreditation body denies an application
for accreditation, the agency shall give notice of the denial, the facts warranting
the denial, and shall afford the applicant an opportunity for a hearing in
accordance with §289.205(h) of this title. If no request for a hearing
is received by the director of the Radiation Control Program within 30 days
of date of receipt of the notice, the agency may proceed to deny. The applicant
shall have the burden of proof showing cause why the application should not
be denied.
(3)
Action on an accreditation application will be abandoned
due to lack of response by the applicant to a request for information by the
agency accreditation body. Abandonment of such actions does not provide an
opportunity for a hearing; however, the applicant retains the right to resubmit
the application and pay a reinstatement fee at any time.
(g)
Suspension and revocation of accreditation of mammography
facilities.
(1)
Suspension of accreditation of mammography facilities.
(A)
An accreditation of a mammography facility may be suspended
or revoked for any of the following reasons:
(i)
any material false statement in the application or any
statement of fact required under provision of the Act;
(ii)
conditions revealed by such application or statement of
fact or any report, record, inspection, or other means that would warrant
the agency accreditation body to refuse to grant an accreditation of mammography
facility on an original application; or
(iii)
failure to observe any of the terms and conditions of
the Act, this chapter, or order of the agency.
(B)
Before the agency accreditation body suspends or revokes
an accreditation of a mammography facility, the agency accreditation body
shall give notice by personal service or by certified mail, addressed to the
last known address, of the facts or conduct alleged to warrant the suspension
or revocation by complaint, and order the accredited mammography facility
to show cause why the mammography facility accreditation should not be suspended
or revoked. The accredited mammography facility shall be given an opportunity
to request a hearing on the matter no later than 30 days after receipt of
the notice.
(C)
Any accredited mammography facility against whom the agency
accreditation body contemplates an action described in subparagraph (A) of
this paragraph may request a hearing by writing the director within 30 days
of receipt of the notice.
(i)
The written request for a hearing must contain the following:
(I)
a statement requesting a hearing; and
(II)
the name, address, and identification number of the accredited
mammography facility against whom the action is being taken.
(ii)
Failure to submit a written request for a hearing within
30 days will render the agency accreditation body action final.
(D)
If the agency accreditation body suspends the accreditation
of a mammography facility in accordance with subparagraph (A) of this paragraph,
the suspension shall remain in effect until the agency accreditation body
determines the following:
(i)
that allegations of violations or misconduct were not substantiated;
(ii)
that violations of required standards have been corrected
to the agency accreditation body's satisfaction; or
(iii)
the facility's accreditation is revoked in accordance
with §289.205 of this title.
(2)
Revocation of accreditation of mammography facilities shall
be in accordance with §289.205(g) of this title.
(h)
Appeal of adverse accreditation or reaccreditation decisions
that preclude certification or recertification.
(1)
The appeal process described in this subsection is available
only for adverse accreditation or reaccreditation decisions that preclude
certification by the agency certifying body. Agency certifying body decisions
to suspend or revoke certificates that are already in effect will be handled
in accordance with §289.230(h) of this title.
(2)
A facility that has been denied accreditation or reaccreditation
is entitled to an appeals process from the agency accreditation body, in accordance
with §289.205 of this title. A facility must avail itself of the accreditation
body's appeal process before requesting reconsideration from the agency certifying
body.
(3)
A facility that cannot achieve satisfactory resolution
of an adverse accreditation decision through the accreditation body's appeal
process is entitled to further appeal to the FDA.
(4)
A facility cannot perform mammography services while an
adverse accreditation decision is being appealed.
(i)
Specific terms and conditions of accreditation of mammography
facilities.
(1)
Each accreditation document issued in accordance with this
section shall be subject to the applicable provisions of the Act, now or hereafter
in effect, and to the applicable requirements and orders of the agency accreditation
body.
(2)
No accreditation document issued by the agency accreditation
body under this section shall be transferred, assigned, or in any manner disposed
of, either voluntarily or involuntarily, to any person.
(j)
Responsibilities of an accredited facility. A facility
shall notify the agency accreditation body of any changes that would render
the information contained in the application inaccurate.
(k)
Expiration and renewal of accreditation of mammography
facilities.
(1)
The accreditation expires at the end of the day in the
month and year stated on the accreditation document.
(2)
An application for renewal of accreditation with the agency
accreditation body shall be filed in accordance with subsection (d) of this
section and with fees in accordance with §289.204 of this title.
(3)
A mammography facility filing an application for renewal
in accordance with subsection (d) of this section and with fees in accordance
with §289.204 of this title before the existing accreditation expires:
(A)
may continue to perform mammography until the expiration
of the accreditation; or
(B)
if the facility receives written authorization from the
agency accreditation body, may continue to perform mammography until the review
process is complete and the accreditation status has been determined by the
agency accreditation body.
(4)
Accreditation for a mammographic facility is valid for
three years from the date of issuance, unless accreditation of the facility
is suspended or revoked prior to such deadline.
(5)
Issuance of renewal of accreditation shall be in accordance
with subsection (e) of this section.
(l)
Complaints. Each facility accredited by the agency accreditation
body shall do the following:
(1)
establish a written procedure for collecting and resolving
consumer complaints;
(2)
maintain a record of each serious complaint received by
the facility in accordance with §289.230(ff)(3) of this title; and
(3)
report unresolved serious complaints to the accreditation
body within 30 days of receiving the complaint.
(m)
Clinical image quality. Clinical images produced by any
certified facility must continue to comply with the standards for clinical
image quality established by that facility's accreditation body.
(n)
Additional mammography review and patient notification.
If the agency certifying body or the agency accreditation body believes that
mammography quality at a facility may have been compromised and presents a
serious risk to human health, the facility shall provide clinical images and
other relevant information, as specified by the agency certifying body or
the agency accreditation body for review by the accreditation body.
(o)
Record requirements. Records required by this section shall
be maintained for inspection by the agency in accordance with subsection §289.230(ff)(3)
of this title. Records may be maintained electronically in accordance with §289.231(ff)(3)
of this title.
(p)
On-site facility visit, targeted clinical image review,
and random clinical image review.
(1)
Each accredited facility shall afford the agency accreditation
body, at all reasonable times, opportunity to audit the facility where mammography
equipment or associated equipment is used or stored.
(2)
Each accredited facility shall make available to the agency
accreditation body for inspection, upon reasonable notice, records maintained
in accordance with this chapter.
(3)
Each accredited facility shall, upon request by the agency
accreditation body or the agency certifying body, make clinical images available
to the image review board for a targeted clinical image review or a random
clinical image review. The agency certifying body, the agency accreditation
body, another FDA-approved accreditation body, or the FDA may request a targeted
clinical image review due to, but not limited to, serious complaints or severe
items of non-compliance.
(4)
Annually, the agency accreditation body shall conduct on-site
visits and random clinical image reviews of a sample of facilities to monitor
and assess their compliance with standards established by the accreditation
body. Other on-site visits may be conducted based on problems identified through
inspections, serious complaints received from consumers or others, a previous
history of noncompliance, or any other information in the possession of the
accreditation body, inspectors, or FDA.
This agency hereby certifies that the proposal has been
reviewed by legal counsel and found to be within the agency's legal authority
to adopt.
Filed with the Office of
the Secretary of State on June 9, 2006.
TRD-200603132
Cathy Campbell
General Counsel
Department of State Health Services
Earliest possible date of adoption: July 23, 2006
For further information, please call: (512) 458-7111 x6972
Chapter 1051.
RULES
Texas Department of Health
].
Pet reptiles
] should be
kept out of households
that include
[
where
] children
aged <5 years or immunocompromised persons [
live
]. Families
expecting a new child should remove
any
[
the pet
] reptile
from the home before the infant arrives.
Pet reptiles
] should not
be
allowed
[
kept
] in childcare centers.
Pet reptiles
] should not
be allowed to roam freely throughout the home or living area.
Pet reptiles
] should be
kept out of kitchens and other food-preparation areas to prevent contamination.
Kitchen sinks should not be used to bathe reptiles or to wash their dishes,
cages, or aquariums.
If bathtubs are used for these purposes, they should
be cleaned thoroughly and disinfected with bleach.
Subchapter G. CAGING REQUIREMENTS AND STANDARDS FOR DANGEROUS WILD ANIMALS
animal
] to a limited amount of space,
including a cage, pen, run, room, compartment, or hutch.
the protection and welfare of
]
the animals
; prevent escape by the animals; and protect and enhance the
public's health and safety
[
and the safety of handlers and the
public
]. Such equipment includes, but is not limited to
:
[
.
]
animal
] from the elements (weather conditions).
Such structures may vary in size depending on the security and biological
needs of the species. The structures are particularly described as follows.
animal
]. When vegetation
and landscaping is available to serve as protection from the elements, access
to a shelter shall also be provided during inclement weather conditions. Such
shelter shall be attached to or adjacent to the primary enclosure.
animal
] from injury and shall be equipped with structural
safety barriers to prevent any public contact with the
animal(s)
[
animal
]. Structural barriers may be constructed from materials such
as fencing, landscaping, or close-mesh wire, provided that materials used
are safe and effective in preventing public contact.
animals are
] housed or exercised
outdoors. Perimeter fences constructed of materials, such as chain link or
welded wire, that allow objects to be passed through them shall be at least
3 feet from the primary enclosure or exercise area.
as
]
that prescribed for cage construction for that particular species. The area
occupied by the safety entrance shall be in addition to the space requirements
for the primary enclosure.
an
] accessible
devices
[
device
] to provide physical stimulation or manipulation
compatible with the species.
Each
[
Such
] device shall
be noninjurious and may include, but is not limited to, boxes, balls, bones,
barrels, drums, rawhide materials, or pools. The area occupied by a pool shall
be in addition to the space requirements for the primary enclosure.
platform
] large enough to accommodate all animals in the
enclosure simultaneously.
a
] claw log.
an
] accessible
devices
[
device
] to provide physical stimulation or manipulation
compatible with the species.
Each
[
Such
] device shall
be noninjurious and may include, but is not limited to, boxes, balls, bones,
barrels, drums, climbing apparatus, or foraging items.
an
] accessible
devices
[
device
] to provide physical stimulation or manipulation
compatible with the species.
Each
[
Such
] device shall
be noninjurious and may include, but is not limited to, boxes, balls, bones,
barrels, drums, rawhide materials, or pools. The area occupied by a pool shall
be in addition to the space requirements for the primary enclosure.
platform
] large enough to accommodate all animals in the
enclosure simultaneously.
platform
] large enough to accommodate all animals in the
enclosure simultaneously.
Chapter 289.
RADIATION CONTROL
Part 15.
COUNCIL ON CARDIOVASCULAR DISEASE AND STROKE